Abstract
Sixty consecutive intertrochanteric femoral fractures were treated with the Asiatic gamma nail followed by early postoperative weight bearing. Cutting-out of the lag screw from the femoral head occurred in six fractures and were analysed radiographically with regard to fracture type (Evans' classification), osteoporosis (Singh grade), accuracy of post-operative reduction (neck-shaft angle and diastasis of the fracture), and location of the lag screw in the femoral head (depth, height, and antero-postero-posterior radiograph and large antero-posterior deviation of the lag screw were significantly related to the increased incidence of the screw cutting-out. This suggests that deep insertion of the lag screw on the antero-posterior view with central placement on the lateral view is an optimal location in the femoral head. The ¿cut-out index' (the multiplication value of the depth and the anteroposterior deviation of the lag screw in the femoral head) is an excellent measure of the risk of the lag screw cutting-out.Entities:
Mesh:
Year: 1998 PMID: 9659482 DOI: 10.1016/s0020-1383(97)00158-7
Source DB: PubMed Journal: Injury ISSN: 0020-1383 Impact factor: 2.586