Evaluation of the VITEK 2 system for rapid identification of medically relevant gram-negative rods.

G Funke1, D Monnet, C deBernardis, A von Graevenitz, J Freney.   

Abstract

The new VITEK 2 system (bioMérieux) was evaluated at two independent sites with the identification card for gram-negative bacilli (ID-GNB card). Of the 845 strains tested, which represented 70 different taxa belonging to either the family Enterobacteriaceae or the nonenteric bacilli, 716 (84.7%) were correctly identified at the species level. Thirty-two (3.8%) additional strains were identified to the species level after the performance of simple, rapid manual tests (oxidase, hemolysis, indole reaction, motility, and pigmentation). For 80 (9.5%) strains, these additional tests did not lead to an identification at the species level but the correct species identification was given among the organisms listed. Only 7 (0.8%) strains were misidentified, and 10 (1.2%) were not identified. Mistakes were randomly distributed over different taxa. Due to the new, more sensitive fluorescence-based technology of the VITEK 2 system, final results were available after 3 h. Since our evaluation was mainly a stress test, it is predicted that the VITEK 2 system in conjunction with the ID-GNB card would perform well under conditions of a routine clinical laboratory in identifying members of the family Enterobacteriaceae and selected species of nonenteric bacteria. This system is a promising, highly automated new tool for the rapid identification of gram-negative bacilli from human clinical specimens.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9650942      PMCID: PMC104958     

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  8 in total

Review 1.  Automated systems for identification of microorganisms.

Authors:  C E Stager; J R Davis
Journal:  Clin Microbiol Rev       Date:  1992-07       Impact factor: 26.132

2.  Evaluating biochemical identification systems.

Authors:  J M Miller
Journal:  J Clin Microbiol       Date:  1991-08       Impact factor: 5.948

3.  Comparison of the autoSCAN-W/A rapid bacterial identification system and the Vitek AutoMicrobic system for identification of gram-negative bacilli.

Authors:  M A Pfaller; D Sahm; C O'Hara; C Ciaglia; M Yu; N Yamane; G Scharnweber; D Rhoden
Journal:  J Clin Microbiol       Date:  1991-07       Impact factor: 5.948

4.  Evaluation of Vitek GNI+ and Becton Dickinson Microbiology Systems Crystal E/NF identification systems for identification of members of the family Enterobacteriaceae and other gram-negative, glucose-fermenting and non-glucose-fermenting bacilli.

Authors:  C M O'Hara; G L Westbrook; J M Miller
Journal:  J Clin Microbiol       Date:  1997-12       Impact factor: 5.948

5.  Evaluation of a small "conventional" identification system for fermentative gramnegative rods using a computerized data base.

Authors:  H H Siegrist; M Altwegg; A Wenger; H Isenrich; A von Graevenitz
Journal:  Zentralbl Bakteriol Mikrobiol Hyg A       Date:  1987-10

6.  Comparison of Crystal Enteric/Nonfermenter system, API 20E system, and Vitek AutoMicrobic system for identification of gram-negative bacilli.

Authors:  A Robinson; Y S McCarter; J Tetreault
Journal:  J Clin Microbiol       Date:  1995-02       Impact factor: 5.948

Review 7.  Biochemical identification of new species and biogroups of Enterobacteriaceae isolated from clinical specimens.

Authors:  J J Farmer; B R Davis; F W Hickman-Brenner; A McWhorter; G P Huntley-Carter; M A Asbury; C Riddle; H G Wathen-Grady; C Elias; G R Fanning
Journal:  J Clin Microbiol       Date:  1985-01       Impact factor: 5.948

8.  Ability of commercial identification systems to identify newly recognized species of Citrobacter.

Authors:  C M O'Hara; S B Roman; J M Miller
Journal:  J Clin Microbiol       Date:  1995-01       Impact factor: 5.948

  8 in total
  46 in total

1.  Evaluation of the Wider system, a new computer-assisted image-processing device for bacterial identification and susceptibility testing.

Authors:  R Cantón; M Pérez-Vázquez; A Oliver; B Sánchez Del Saz; M O Gutiérrez; M Martínez-Ferrer; F Baquero
Journal:  J Clin Microbiol       Date:  2000-04       Impact factor: 5.948

2.  Evaluation of VITEK 2 rapid identification and susceptibility testing system against gram-negative clinical isolates.

Authors:  T K Ling; P C Tam; Z K Liu; A F Cheng
Journal:  J Clin Microbiol       Date:  2001-08       Impact factor: 5.948

3.  Comparative evaluation of the BD Phoenix and VITEK 2 automated instruments for identification of isolates of the Burkholderia cepacia complex.

Authors:  Sylvain Brisse; Stefania Stefani; Jan Verhoef; Alex Van Belkum; Peter Vandamme; Wil Goessens
Journal:  J Clin Microbiol       Date:  2002-05       Impact factor: 5.948

4.  Evaluation of a new system, VITEK 2, for identification and antimicrobial susceptibility testing of enterococci.

Authors:  F Garcia-Garrote; E Cercenado; E Bouza
Journal:  J Clin Microbiol       Date:  2000-06       Impact factor: 5.948

5.  Accuracy of six commercially available systems for identification of members of the family vibrionaceae.

Authors:  Caroline M O'Hara; Evangeline G Sowers; Cheryl A Bopp; Sarah B Duda; Nancy A Strockbine
Journal:  J Clin Microbiol       Date:  2003-12       Impact factor: 5.948

6.  Evaluation of the Vitek 2 system for rapid identification of clinical isolates of gram-negative bacilli and members of the family Streptococcaceae.

Authors:  P J Gavin; J R Warren; A A Obias; S M Collins; L R Peterson
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2002-12-10       Impact factor: 3.267

7.  Evaluation of the VITEK 2 system for rapid direct identification and susceptibility testing of gram-negative bacilli from positive blood cultures.

Authors:  Thomas K W Ling; Z K Liu; Augustine F B Cheng
Journal:  J Clin Microbiol       Date:  2003-10       Impact factor: 5.948

8.  Use of positive blood cultures for direct identification and susceptibility testing with the vitek 2 system.

Authors:  Marina de Cueto; Esther Ceballos; Luis Martinez-Martinez; Evelio J Perea; Alvaro Pascual
Journal:  J Clin Microbiol       Date:  2004-08       Impact factor: 5.948

9.  Evaluation of the new VITEK 2 card for identification of clinically relevant gram-negative rods.

Authors:  Guido Funke; Pascale Funke-Kissling
Journal:  J Clin Microbiol       Date:  2004-09       Impact factor: 5.948

Review 10.  Methylobacterium and its role in health care-associated infection.

Authors:  Julia Kovaleva; John E Degener; Henny C van der Mei
Journal:  J Clin Microbiol       Date:  2014-01-15       Impact factor: 5.948

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.