Literature DB >> 9648088

The acu-men: a new device for continuous renal replacement therapy in acute renal failure.

M Manns1, H D Polaschegg, C Schlaeper, B Steinbach, H G Evering.   

Abstract

BACKGROUND: We sought to design a simple machine to safely provide continuous veno-venous hemofiltration to acute renal failure patients.
RESULTS: The acu-men device uses a pneumatic blood pump with tidal blood flow as the driving force. A volumetric balancing system balances the filtrate with the replacement fluid, and the blood-air interface is eliminated by replacing the conventional venous drip chamber with two air-separating membranes. The extracorporeal circuit is integrated in a disposable cartridge, which is inserted into the machine at the beginning of treatment. The priming and rinsing is done automatically.
CONCLUSION: While preliminary data from an ongoing clinic trial on the efficacy of the device are encouraging, further long-term studies are necessary to evaluate its potential to decrease morbidity and mortality in acute renal failure patients.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9648088     DOI: 10.1046/j.1523-1755.1998.00959.x

Source DB:  PubMed          Journal:  Kidney Int        ISSN: 0085-2538            Impact factor:   10.612


  2 in total

1.  Technical breakthroughs in the wearable artificial kidney (WAK).

Authors:  Victor Gura; Alexandra S Macy; Masoud Beizai; Carlos Ezon; Thomas A Golper
Journal:  Clin J Am Soc Nephrol       Date:  2009-08-20       Impact factor: 8.237

2.  Development of an accurate fluid management system for a pediatric continuous renal replacement therapy device.

Authors:  Arvind Santhanakrishnan; Trent T Nestle; Brian L Moore; Ajit P Yoganathan; Matthew L Paden
Journal:  ASAIO J       Date:  2013 May-Jun       Impact factor: 2.872

  2 in total