Mobility assessment with the Periotest system in relation to histologic findings of oral implants.

F Isidor1.   

Abstract

The relationship between mobility assessment with the Periotest system and histologic findings was evaluated for oral implants. Five screw-type implants of pure titanium were placed in the mandibles of four monkeys. Two implants in each monkey were occlusally overloaded. These implants were brushed once a week. Plaque was allowed to accumulate around unloaded implants with abutments in the same monkeys. During the experiment, six of eight implants with occlusal overload showed increased manually detectable mobility. Two of these were lost. After 18 months of experimentation, the mobility was assessed using the Periotest system. Sections of the implants and surrounding tissue were cut. For the excessive occlusally loaded implants with manually detectable mobility, positive Periotest values were recorded, and for all other implants the values were negative (range = -7 to -2). All implants with plaque accumulation were histologically osseointegrated but showed marginal bone loss. Two of the implants with occlusal overload had lost osseointegration completely, and two other implants were osseointegrated in the apical part only. A statistically significant association between the Periotest values and the histologic bone level or the proportion of bone-implant contact was observed. If only clinically stable implants (i.e., without manually detectable mobility or with a negative Periotest value) were included in the analysis, no significant correlation was found. The Periotest values revealed only slightly more information concerning the osseointegration of implants than manual mobility assessments.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9638008

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  7 in total

1.  Bone response to immediate loading through titanium implants with different surface roughness in rats.

Authors:  Naoko Sato; Toshie Kuwana; Miou Yamamoto; Hanako Suenaga; Takahisa Anada; Shigeto Koyama; Osamu Suzuki; Keiichi Sasaki
Journal:  Odontology       Date:  2013-04-07       Impact factor: 2.634

2.  Influence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study.

Authors:  David Chávarri-Prado; Aritza Brizuela-Velasco; Markel Diéguez-Pereira; Esteban Pérez-Pevida; Antonio Jiménez-Garrudo; Iratxe Viteri-Agustín; Alejandro Estrada-Martínez; Oier Montalbán-Vadillo
Journal:  J Clin Exp Dent       Date:  2020-03-01

3.  The effect of the thread depth on the mechanical properties of the dental implant.

Authors:  Sun-Young Lee; Sung-Jun Kim; Hyun-Wook An; Hyun-Seung Kim; Dong-Guk Ha; Kyung-Ho Ryo; Kwang-Bum Park
Journal:  J Adv Prosthodont       Date:  2015-04-23       Impact factor: 1.904

4.  Influence of occlusal loading on peri-implant clinical parameters. A pilot study.

Authors:  Hilario Pellicer-Chover; José Viña-Almunia; Javier Romero-Millán; David Peñarrocha-Oltra; Berta García-Mira; María Peñarrocha-Diago
Journal:  Med Oral Patol Oral Cir Bucal       Date:  2014-05-01

5.  Fabrication of a screw-retained hybrid prosthesis following treatment of peri-implant defects: a case report.

Authors:  Bebek Serra Oguz Ahmet; Gulsum Sayin Ozel; Hilal Uslu Toygar
Journal:  J Istanb Univ Fac Dent       Date:  2016-10-01

Review 6.  Clinical Assessment of Dental Implant Stability During Follow-Up: What Is Actually Measured, and Perspectives.

Authors:  Elisabetta M Zanetti; Giulia Pascoletti; Michele Calì; Cristina Bignardi; Giordano Franceschini
Journal:  Biosensors (Basel)       Date:  2018-07-13

7.  Oral status and aesthetics after nonsurgical periodontal treatment: Do patient's perception and dentist's evaluation agree?

Authors:  Manuela Elena Kaufmann; Deborah Hofer; Daniel B Wiedemeier; Thomas Attin; Patrick R Schmidlin
Journal:  Clin Exp Dent Res       Date:  2019-08-19
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.