M Marsh1, O Tu, V Dolnik, D Roach, N Solomon, K Bechtol, P Smietana, L Wang, X Li, P Cartwright, A Marks, D Barker, D Harris, J Bashkin.
Abstract
A capillary array electrophoresis apparatus capable of running and analyzing 48 DNA sequencing samples simultaneously has been constructed. The instrument uses a replaceable sieving buffer and incorporates a convenient method for introducing the buffer into the capillaries. Data from laser-induced fluorescence are collected as four separate images, one for each optical channel. The integrated data analysis software employs an open architecture that allows use of any DNA base-calling algorithm. DNA sequencing runs are completed in approx. 1 hr (approximately 500 bases), and instrument turnaround time between runs is less than 15 min. Overall, the instrument throughput is on the order of 720 templates/day, or 360,000 bases/day.Mesh:
Year: 1997 PMID: 9624574
Source DB: PubMed Journal: J Capillary Electrophor ISSN: 1079-5383