Prolonged prothrombin time and activated partial thromboplastin time due to underfilled specimen tubes with 109 mmol/L (3.2%) citrate anticoagulant.

J Reneke1, J Etzell, S Leslie, V L Ng, E L Gottfried.   

Abstract

Underfilling of specimen tubes containing 129 mmol/L (3.8%) buffered citrate prolongs prothrombin time (PT) and activated partial thromboplastin time (APTT) values. We studied this phenomenon by using 109 mmol/L (3.2%) buffered citrate as the anticoagulant, anticipating some increase in tolerance to underfilling. Venous blood drawn from 12 healthy subjects and 30 patients receiving long-term oral warfarin therapy was mixed with 109 mmol/L buffered citrate solution in proportions equivalent to filling the collection tubes from 52% to 100% of capacity. Accurate PT values were obtained from normal specimens if the tubes were filled to 65% or more of capacity. Accurate PT results in the therapeutic range were obtained only with filling to 80% or more of capacity (using a "moderately sensitive" thromboplastin reagent, International Sensitivity Index [ISI] = 2.06) or 90% or more of capacity (using a "highly sensitive" thromboplastin reagent, ISI = 1.01). In contrast, APTT was much less tolerant to underfilling, with prolonged values observed in most specimens filled to less than 90% of capacity. No false low values were observed. Specimen tubes should be filled to at least 90% of capacity to avoid falsely elevated PT or APTT results, but values within the reference range may be acceptable even from underfilled tubes.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9620035     DOI: 10.1093/ajcp/109.6.754

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  4 in total

1.  Influence of blood specimen collection method on various preanalytical sample quality indicators.

Authors:  Tester F Ashavaid; Sucheta P Dandekar; Bhamini Keny; Vishaal R Bhambhwani
Journal:  Indian J Clin Biochem       Date:  2008-06-11

2.  Quantitative assessment of prevalence of pre-analytical variables and their effect on coagulation assay. Can intervention improve patient safety?

Authors:  Ravi Bhushan; Arijit Sen
Journal:  Med J Armed Forces India       Date:  2017-01-05

3.  Effect of anticoagulant adjustment on prothrombin time test using two different PT reagents in patients with elevated hematocrit.

Authors:  Vanessa M Silva; Denise C Rezende; Edna S Garcia; Cyrillo Cavalheiro; Celia C Strunz
Journal:  Pract Lab Med       Date:  2020-10-17

4.  Influence of blood collection systems on coagulation tests.

Authors:  Soner Yavaş; Selime Ayaz; S Kenan Köse; Fatma Ulus; A Tulga Ulus
Journal:  Turk J Haematol       Date:  2012-12-05       Impact factor: 1.831

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.