Literature DB >> 9609877

Yasargil aneurysm clips: evaluation of interactions with a 1.5-T MR system.

F G Shellock1, E Kanal.   

Abstract

PURPOSE: To evaluate the magnetic field interactions with Yasargil aneurysm clips exposed to a 1.5-T MR system.
MATERIALS AND METHODS: Sixteen Yasargil aneurysm clips made with Phynox (10 clips returned to the manufacturer due to suspected magnetic field interactions and six clips obtained from the manufacturer in sterile packages) were tested for magnetic field interaction at 1.5 T by using techniques to assess magnetic field-induced translational (deflection angle test) and torque (changes in alignment or rotation) forces.
RESULTS: None of the aneurysm clips demonstrated magnetic field interactions; that is, deflection angles were zero and there were no changes in alignment or evidence of rotational forces present.
CONCLUSION: It is safe to perform MR procedures at 1.5 T or less in patients with Yasargil aneurysm clips made with nonferromagnetic material.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9609877     DOI: 10.1148/radiology.207.3.9609877

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  3 in total

1.  Artifact quantification and tractography from 3T MRI after placement of aneurysm clips in subarachnoid hemorrhage patients.

Authors:  Faraz Khursheed; Fiona Rohlffs; Shuichi Suzuki; Dong H Kim; Timothy M Ellmore
Journal:  BMC Med Imaging       Date:  2011-10-04       Impact factor: 1.930

2.  Aneurysm clips: evaluation of magnetic field interactions and translational attraction by use of "long-bore" and "short-bore" 3.0-T MR imaging systems.

Authors:  Frank G Shellock; Jean A Tkach; Paul M Ruggieri; Thomas J Masaryk; Peter A Rasmussen
Journal:  AJNR Am J Neuroradiol       Date:  2003-03       Impact factor: 3.825

3.  In vitro evaluation of MR imaging issues at 3T for aneurysm clips made from MP35N: Findings and information applied to 155 additional aneurysm clips.

Authors:  F G Shellock; S Valencerina
Journal:  AJNR Am J Neuroradiol       Date:  2009-12-24       Impact factor: 3.825

  3 in total