Literature DB >> 9607367

Repositioning accuracy: comparison of a noninvasive head holder with thermoplastic mask for fractionated radiotherapy and a case report.

R Sweeney1, R Bale, M Vogele, M Nevinny-Stickel, A Bluhm, T Auer, G Hessenberger, P Lukas.   

Abstract

PURPOSE: To compare accuracy, clinical feasibility, and subjective patient impression between a noninvasive head holder (Vogele Bale Hohner [VBH]; Wellhoefer Dosimetry, Schwarzenbruck, Germany) developed at the University of Innsbruck and the thermoplastic mask fixation system for use in fractionated external radiotherapy. We present a case report of an actual patient fixated in the VBH head holder during radiation therapy.
MATERIALS AND METHODS: The VBH head holder consists of an individualized vacuum dental cast connected to a head plate via two hydraulic arms allowing noninvasive, reproducible head fixation of even uncooperative patients. Accuracy was tested and compared with that of the thermoplastic mask using the Phillips EasyGuide navigation system on five volunteers. Specific external registration points served as landmarks and their positions were compared after each repositioning. System and operator inaccuracy were also taken into account. The times taken for production and repositioning of the respective fixation devices were compared, and subjective impressions were noted.
RESULTS: Mean VBH head holder repositioning accuracy was 1.02 mm while that of the thermoplastic mask was 3.05 mm. 69% of mask repositionings showed a deviation > 2 mm and 41% > 3 mm (as opposed to 8% and 1% respectively for the VBH head holder) Those points located farthest away from the respective plane of fixation showed the largest deviations. Both production and repositioning times were similar between the systems; depending upon the patient, the VBH head holder was generally better tolerated than the mask system.
CONCLUSION: Due to its significantly better repositioning accuracy compared to that of the thermoplastic mask, the VBH head holder is especially suited for external radiation requiring precise repositioning due to critical tissues in immediate surrounding of the area to be irradiated.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9607367     DOI: 10.1016/s0360-3016(98)00064-9

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  13 in total

1.  Radiosurgery technology development and use.

Authors:  Sanford L Meeks; Jason Pukala; Naren Ramakrishna; Twyla R Willoughby; Francis J Bova
Journal:  J Radiosurg SBRT       Date:  2011

2.  Development of a Real-Time Thermoplastic Mask Compression Force Monitoring System Using Capacitive Force Sensor.

Authors:  Tae-Ho Kim; Min-Seok Cho; Dong-Seok Shin; Dong Ho Shin; Siyong Kim
Journal:  Front Robot AI       Date:  2022-07-06

3.  Robot-assisted mediastinal parathyroidectomy.

Authors:  C Profanter; T Schmid; R Prommegger; R Bale; T Sauper; J Bodner
Journal:  Surg Endosc       Date:  2004-02-02       Impact factor: 4.584

4.  Semi-robotic 6 degree of freedom positioning for intracranial high precision radiotherapy; first phantom and clinical results.

Authors:  Jürgen Wilbert; Matthias Guckenberger; Bülent Polat; Otto Sauer; Michael Vogele; Michael Flentje; Reinhart A Sweeney
Journal:  Radiat Oncol       Date:  2010-05-26       Impact factor: 3.481

5.  Accurate positioning for head and neck cancer patients using 2D and 3D image guidance.

Authors:  Hyejoo Kang; Dale M Lovelock; Ellen D Yorke; Sergey Kriminski; Nancy Lee; Howard I Amols
Journal:  J Appl Clin Med Phys       Date:  2010-10-27       Impact factor: 2.102

6.  [Noninvasive head fixation for external irradiation of tumors of the head-neck area].

Authors:  R J Bale; R Sweeney; M Vogele; M Nevinny; T Auer; A Bluhm; W F Thumfart; P Lukas
Journal:  Strahlenther Onkol       Date:  1998-07       Impact factor: 3.621

7.  Virtual neck exploration in patients with hyperparathyroidism and former cervical operations.

Authors:  Gerd Wimmer; Reto Bale; Peter Kovacs; Michael Gabriel; Daniel Putzer; Tonja Sauper; Michael Sieb; Christoph Profanter; Raimund Margreiter; Rupert Prommegger
Journal:  Langenbecks Arch Surg       Date:  2008-06-25       Impact factor: 3.445

8.  Intra-fraction motion gating during frameless Gamma Knife® Icon™ therapy: The relationship between cone beam CT assessed intracranial anatomy displacement and infrared-tracked nose marker displacement.

Authors:  Gavin Wright; Jannie Schasfoort; Natalie Harrold; Paul Hatfield; Peter Bownes
Journal:  J Radiosurg SBRT       Date:  2019

9.  Analysis of inter- and intrafraction accuracy of a commercial thermoplastic mask system used for image-guided particle radiation therapy.

Authors:  Dante Amelio; Marcus Winter; Daniel Habermehl; Oliver Jäkel; Jurgen Debus; Stephanie E Combs
Journal:  J Radiat Res       Date:  2013-07       Impact factor: 2.724

10.  Preliminary Evaluation of a Novel Thermoplastic Mask System with Intra-fraction Motion Monitoring for Future Use with Image-Guided Gamma Knife.

Authors:  Winnie Li; Gregory Bootsma; Oscar Von Schultz; Per Carlsson; Normand Laperriere; Barbara-Ann Millar; David Jaffray; Caroline Chung
Journal:  Cureus       Date:  2016-03-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.