EBRA improves the accuracy of radiographic analysis of acetabular cup migration.

T Ilchmann1, U Kesteris, H Wingstrand.   

Abstract

EBRA (Ein Bild Röntgen Analyse) is a new computerized method measuring migration and wear of the acetabular cup, suggested to improve measurement accuracy. We evaluated possible errors of measurement and compared EBRA with standard methods. 1. We did repeated measurements on a single radiograph using the same reference lines. The reliability of the input procedure with standard measurements was significantly better than repeated digitization with EBRA. 2. In a more clinical test, a group of 10 patients was studied. 5 radiographs were taken of the same patient on the same day. EBRA improved the reliability of repeated radiographic examination significantly for migration measurements in the vertical direction. 3. To assess the inter- and intraobserver variations, repeated measurements were performed on the clinical series of pelvic radiographs of 10 patients. EBRA was significantly better than standard methods. With EBRA, errors of wear and migration measurements could be reduced, as compared to standard methods. The major improvement with EBRA was found for migration measurements in the vertical direction.

Entities:  

Mesh:

Year:  1998        PMID: 9602766     DOI: 10.3109/17453679809117610

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  5 in total

Review 1.  Radiographic assessment of the cup orientation after total hip arthroplasty: a literature review.

Authors:  Jing-Xin Zhao; Xiu-Yun Su; Zhe Zhao; Ruo-Xiu Xiao; Li-Cheng Zhang; Pei-Fu Tang
Journal:  Ann Transl Med       Date:  2020-02

2.  Clinical evaluation of a novel press-fit acetabular cup using "Ein-Bild-Roentgen-Analysis" (EBRA): A positive short-term prognosis.

Authors:  B A Ishaque; J Wieczorek; C A Fonseca Ulloa; J B Seeger; G A Ahmed; M Rickert; A Jahnke
Journal:  J Orthop       Date:  2020-03-30

3.  Migration pattern of cementless press fit cups in the presence of stabilizing screws in total hip arthroplasty.

Authors:  C Zilkens; S Djalali; B Bittersohl; T Kälicke; C N Kraft; R Krauspe; Marcus Jäger
Journal:  Eur J Med Res       Date:  2011-03-28       Impact factor: 2.175

4.  Good mid-term results with the trident peripheral self-locking cup: a clinical evaluation and migration measurement with EBRA.

Authors:  Dietmar Dammerer; Philipp Blum; David Putzer; Andreas Tscholl; Michael C Liebensteiner; Martin Thaler
Journal:  Arch Orthop Trauma Surg       Date:  2020-11-09       Impact factor: 3.067

Review 5.  Temporal subsidence patterns of cemented polished taper-slip stems: a systematic review.

Authors:  Kwaku Baryeh; Jeewaka Mendis; David H Sochart
Journal:  EFORT Open Rev       Date:  2021-05-04
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.