Literature DB >> 9581951

Effect of shock polarity on biphasic defibrillation thresholds using an active pectoral lead system.

M R Olsovsky1, S R Shorofsky, M R Gold.   

Abstract

INTRODUCTION: The downsizing of implantable defibrillator pulse generators has made pectoral placement routine. A further reduction of defibrillation thresholds (DFTs) may simplify implantation defibrillation testing and allow for smaller, lower output pulse generators while maintaining an adequate defibrillation safety margin. One factor that may affect defibrillation efficacy is shock polarity. METHODS AND
RESULTS: Sixty consecutive patients undergoing dual-coil, active left pectoral defibrillator implantation were evaluated. Paired, biphasic DFTs were measured in normal (RV apex = cathode) and reverse (RV apex = anode) polarity with order of testing randomized. Reverse polarity conferred a 15% reduction of mean DFTs (8.5 +/- 5.0 J normal, 7.2 +/- 4.6 J reverse polarity, P = 0.02). The effect of polarity appeared most pronounced among the patients with a high DFT (> or = 15 J) resulting in a 31% reduction with reverse polarity (16.7 +/- 2.5 J normal, 11.5 +/- 5.9 J reverse, P = 0.03).
CONCLUSION: Reversing shock polarity results in significantly lower biphasic DFTs with an active pectoral lead system, particularly in the subgroup of patients with a high normal polarity threshold. Reversing polarity in these patients may simplify acute defibrillation testing and allow for lower output devices.

Entities:  

Mesh:

Year:  1998        PMID: 9581951     DOI: 10.1111/j.1540-8167.1998.tb00923.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  5 in total

1.  Subpectoral cardioverter-defibrillator implantation using a lateral approach.

Authors:  X F Costeas; P G Strembelas; D X Markou; C I Stefanadis; P K Toutouzas
Journal:  J Interv Card Electrophysiol       Date:  2000-12       Impact factor: 1.900

2.  Multicentre comparison Of shock efficacy using single-vs. Dual-coil lead systems and Anodal vs. cathodaL polarITY defibrillation in patients undergoing transvenous cardioverter-defibrillator implantation. The MODALITY study.

Authors:  Maria Stella Baccillieri; Gianni Gasparini; Luca Benacchio; Alessandro Zorzi; Elena Marras; Francesca Zerbo; Luca Tomasi; Diego Vaccari; Gianni Pastore; Carlo Bonanno; Giulio Molon; Gabriele Zanotto; Antonio Fusco; Massimo Carasi; Andrea Zorzi; Vittorio Calzolari; Barbara Ignatiuk; Sergio Cannas; Alessandro Vaglio; Muhamad Al Bunni; Antonella Pedrini; Armando Olivieri; Roberta Rampazzo; Nadia Minicuci; Domenico Corrado; Roberto Verlato
Journal:  J Interv Card Electrophysiol       Date:  2015-02-19       Impact factor: 1.900

3.  A randomized comparison of defibrillation thresholds in the right ventricular outflow tract versus right ventricular apex.

Authors:  Michael Mollerus; Margaret Lipinski; Thomas Munger
Journal:  J Interv Card Electrophysiol       Date:  2008-05-10       Impact factor: 1.900

4.  Effect of shock polarity on defibrillation thresholds with a hybrid patch-coil lead system.

Authors:  Eric J Rashba; Stephen R Shorofsky; Robert W Peters; Michael R Gold
Journal:  J Interv Card Electrophysiol       Date:  2003-12       Impact factor: 1.900

5.  Tetanizing prepulse: A novel strategy to mitigate implantable cardioverter-defibrillator shock-related pain.

Authors:  David W Hunter; Harikrishna Tandri; Henry Halperin; Leslie Tung; Ronald D Berger
Journal:  Heart Rhythm       Date:  2016-01-06       Impact factor: 6.343

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.