Literature DB >> 9555789

Successful percutaneous extraction of pacemaker leads with a novel (VascoExtor) pacing lead removal system.

A S Manolis1, T N Maounis, J Chiladakis, V Vassilikos, H Melita-Manolis, D V Cokkinos.   

Abstract

A new pacing lead extraction system (VascoExtor, VascoMed, Germany) with a universally applicable locking stylet was used over a period of 12 months to extract 25 permanent pacemaker leads (mean implantation time 3.9 +/- 3.8 years) in 16 patients who had pacemaker infection (n = 13), lead dysfunction (n = 2), or before an automatic defibrillator implant (n = 1). With use of this system, we were able to successfully and safely remove 24 of 25 chronic pacing leads (96%) in 15 of 16 patients (93%) with sole use of the locking stylet in 81%, and with use of an array of ancillary tools in the remaining 19% of patients.

Entities:  

Mesh:

Year:  1998        PMID: 9555789     DOI: 10.1016/s0002-9149(98)00026-5

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  6 in total

1.  Percutaneous extraction of transvenous defibrillator leads using the VascoExtor pacing lead removal system.

Authors:  V P Vassilikos; T N Maounis; J Chiladakis; D V Cokkinos; A S Manolis
Journal:  J Interv Card Electrophysiol       Date:  1999-10       Impact factor: 1.900

Review 2.  Endovascular extraction techniques for pacemaker and ICD lead extraction: Part 1.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-04       Impact factor: 2.380

Review 3.  Micafungin for Candida albicans pacemaker-associated endocarditis: a case report and review of the literature.

Authors:  Carlo Tascini; Maria Grazia Bongiorni; Enrico Tagliaferri; Antonello Di Paolo; Sarah Flammini; Ezio Soldati; Alessandro Leonildi; Andrea Di Cori; Francesco Menichetti
Journal:  Mycopathologia       Date:  2012-10-17       Impact factor: 2.574

4.  The lead extractor's toolbox: a review of current endovascular pacemaker and ICD lead extraction techniques.

Authors:  F A Bracke
Journal:  Indian Pacing Electrophysiol J       Date:  2003-07-01

5.  Cardiac implantable electronic device lead extraction using the lead-locking device system: keeping it simple, safe, and inexpensive with mechanical tools and local anesthesia.

Authors:  Antonis S Manolis; Georgios Georgiopoulos; Sofia Metaxa; Spyridon Koulouris; Dimitris Tsiachris
Journal:  Anatol J Cardiol       Date:  2017-08-11       Impact factor: 1.596

6.  Percutaneous extraction of transvenous permanent pacemaker/defibrillator leads.

Authors:  Stylianos Paraskevaidis; Dimitrios Konstantinou; Vassilios Vassilikos; Efstratios Theofilogiannakos; Lilian Mantziari; Athanasia Megarisiotou; Ioannis Galitsianos; Charalambos Karvounis
Journal:  Biomed Res Int       Date:  2014-05-26       Impact factor: 3.411

  6 in total