On the necessity of the invasive predischarge test after implantation of a cardioverter-defibrillator.

A Wieckhorst1, A Buchwald, C Unterberg.   

Abstract

The necessity of routine invasive implantable cardioverter-defibrillator (ICD) testing before hospital discharge was analyzed in 268 patients. In 98% of the patients, invasive ICD testing was not necessary if a preimplant electrophysiologic test was performed; most of the observed problems can be solved by noninvasive control.

Entities:  

Mesh:

Year:  1998        PMID: 9555788     DOI: 10.1016/s0002-9149(98)00025-3

Source DB:  PubMed          Journal:  Am J Cardiol        ISSN: 0002-9149            Impact factor:   2.778


  2 in total

1.  Indications for predismissal testing with arrhythmia-induction in patients receiving an implantable cardioverter defibrillator.

Authors:  G Christ; R Becker; F Voss; K Kelemen; J Senges-Becker; M Hauck; W Schoels; I Bald; H A Katus; A Bauer
Journal:  Clin Res Cardiol       Date:  2007-06-27       Impact factor: 5.460

2.  Defibrillation testing of the implantable cardioverter defibrillator: when, how, and by whom?

Authors:  Luis A Pires
Journal:  Indian Pacing Electrophysiol J       Date:  2007-08-01