Influence of muscle forces on loads in internal spinal fixation devices.

A Rohlmann1, G Bergmann, F Graichen, H M Mayer.   

Abstract

STUDY
DESIGN: The loads acting on an internal spinal fixation device were measured in vivo.
OBJECTIVES: To determine the influence of muscle forces on implant loads. SUMMARY OF BACKGROUND DATA: Only limited information exists regarding the loads acting on spinal implants in vivo. Though the muscles greatly influence spinal load, they have been neglected in most studies.
METHODS: Telemeterized internal spinal fixation devices were used to study the influence of muscle forces on the implant loads in three patients before and after anterior interbody fusion.
RESULTS: Contracting abdominal or back muscles in a lying position was found to significantly increase implant loads. Hanging by the hands from wall bars as well as balancing with the hands on parallel bars reduced the implant loads compared with standing; however, hanging by the feet with the head upside down did not reduce implant loads compared with lying in a supine position. When lying on an operating table with only the foot end lowered so that the hips were bent, the patient had different load measurements in the conscious and anesthetized state before anterior interbody fusion. The anesthetized patient evidenced predominately extension moments in both fixators, whereas flexion moments were observed in the right fixator of the conscious patient. After anterior interbody fusion had occurred, the differences in implant loads resulting from anesthesia were small.
CONCLUSIONS: The muscles greatly influence implant loads. They prevent an axial tensile load on the spine when part of the body weight is pulling, e.g., when the patient is hanging by his hands or feet. The implant loads may be strongly altered when the patient is under anesthesia.

Entities:  

Mesh:

Year:  1998        PMID: 9530784     DOI: 10.1097/00007632-199803010-00005

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  5 in total

Review 1.  Parameters that effect spine biomechanics following cervical disc replacement.

Authors:  Vijay K Goel; Ahmad Faizan; Vivek Palepu; Sanghita Bhattacharya
Journal:  Eur Spine J       Date:  2011-05-20       Impact factor: 3.134

Review 2.  'SMART' implantable devices for spinal implants: a systematic review on current and future trends.

Authors:  Sihyong J Kim; Tian Wang; Matthew H Pelletier; William R Walsh
Journal:  J Spine Surg       Date:  2022-03

3.  Development and Biomechanical Study of a New Open Dynamic Anterior Cervical Nail Plate System.

Authors:  Xiao-Feng Zhao; Yi-Bo Zhao; Xiang-Dong Lu; Wen-Xuan Wang; De-Tai Qi; Xu Yang; Xiao-Nan Wang; Run-Tian Zhou; Yuan-Zhang Jin; Bin Zhao
Journal:  Orthop Surg       Date:  2020-02       Impact factor: 2.071

4.  Continuous Rod Load Monitoring to Assess Spinal Fusion Status-Pilot In Vivo Data in Sheep.

Authors:  Markus Windolf; Maximilian Heumann; Viktor Varjas; Caroline Constant; Manuela Ernst; Robert Geoff Richards; Hans-Joachim Wilke; Lorin Michael Benneker
Journal:  Medicina (Kaunas)       Date:  2022-07-06       Impact factor: 2.948

5.  Smart implants in orthopedic surgery, improving patient outcomes: a review.

Authors:  Eric H Ledet; Benjamin Liddle; Katerina Kradinova; Sara Harper
Journal:  Innov Entrep Health       Date:  2018-08-29
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.