Potential renal transplant donors: evaluation with gadolinium-enhanced MR angiography and MR urography.

R N Low1, A G Martinez, S M Steinberg, G D Alzate, K E Kortman, B B Bower, W J Dwyer, S K Prince.   

Abstract

PURPOSE: To determine prospectively the feasibility and accuracy of combined gadolinium-enhanced magnetic resonance (MR) angiography, MR urography, and MR nephrography in the presurgical evaluation of potential renal transplant donors.
MATERIALS AND METHODS: Twenty-two potential donors for renal transplantation were evaluated with 1.5-T MR imaging. MR angiograms were evaluated for the number of renal arteries, presence of early arterial branches, and renal artery stenoses. The renal collecting system and ureters were evaluated on the MR urograms. Renal parenchyma was assessed on the MR nephrogram. Prospective interpretation of MR images was compared with that of conventional angiograms and excretory urograms and with surgical findings.
RESULTS: Gadolinium-enhanced MR angiography enabled correct identification of the arterial supply to all 44 native kidneys (44 single or dominant renal arteries and nine accessory renal arteries), four of five early arterial branches arising in the proximal 2 cm of the renal artery, a mild truncal stenosis in one renal artery, and two anomalies of the draining renal veins. The MR urogram accurately depicted a duplicated collecting system and mild unilateral pelvicalicectasis. The MR nephrogram showed renal size and a solitary cyst in one kidney, confirmed with sonography.
CONCLUSION: Combined gadolinium-enhanced MR angiography, MR urography, and MR nephrography can accurately depict the arterial supply, collecting system, and renal parenchyma of donor kidneys.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9530313     DOI: 10.1148/radiology.207.1.9530313

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  5 in total

1.  Evaluation of renal arteries in living renal donors: comparison between MDCT angiography and gadolinium-enhanced 3D MR angiography.

Authors:  Tonsok Kim; Takamichi Murakami; Satoru Takahashi; Masatoshi Hori; Shirou Takahara; Naotsugu Ichimaru; Akihiko Okuyama; Yoshifumi Narumi; Hironobu Nakamura
Journal:  Radiat Med       Date:  2006-11-24

2.  An automatic method for renal cortex segmentation on CT images: evaluation on kidney donors.

Authors:  Xinjian Chen; Ronald M Summers; Monique Cho; Ulas Bagci; Jianhua Yao
Journal:  Acad Radiol       Date:  2012-02-15       Impact factor: 3.173

Review 3.  [MR urography: principles, examination techniques, indications].

Authors:  M Memarsadeghi; M Riccabona; G Heinz-Peer
Journal:  Radiologe       Date:  2005-10       Impact factor: 0.635

4.  Magnetic resonance urography for the assessment of potential renal donors: comparison of the RARE technique with a low-dose gadolinium-enhanced magnetic resonance urography technique in the absence of pharmacological and mechanical intervention.

Authors:  Klaus D Hagspiel; Sabah Butty; Kiran R Nandalur; Eric A Bissonette; Ming-Chen Paul Shih; Daniel A Leung; J Fritz Angle; David J Spinosa; Alan H Matsumoto; Hossam Ahmed; Hilary Sanfey; Ross B Isaacs; Robert G Sawyer; Timothy L Pruett
Journal:  Eur Radiol       Date:  2005-07-14       Impact factor: 5.315

5.  MR urography for suspected upper tract urothelial carcinoma.

Authors:  Naoki Takahashi; Akira Kawashima; James F Glockner; Robert P Hartman; Bohyun Kim; Bernard F King
Journal:  Eur Radiol       Date:  2008-11-27       Impact factor: 5.315

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.