Literature DB >> 9525563

Catheter closure of moderate- to large-sized patent ductus arteriosus using the new Amplatzer duct occluder: immediate and short-term results.

J Masura1, K P Walsh, B Thanopoulous, C Chan, J Bass, Y Goussous, P Gavora, Z M Hijazi.   

Abstract

OBJECTIVES: The aim of this study was to assess the immediate and short-term results of anterograde catheter closure of a moderate- to large-sized patent ductus arteriosus (PDA) using the new self-expandable, respositionable Amplatzer duct occluder (ADO) device.
BACKGROUND: Transcatheter closure of a PDA using devices or coils is technically challenging and may be accompanied by a 38% incidence of residual shunts.
METHODS: Twenty-four patients (6 male, 18 female) underwent attempted transcatheter closure of a PDA using the ADO at a median age of 3.8 years (range 0.4 to 48) and a median weight of 15.5 kg (range 6 to 70). The mean PDA diameter at its narrowest segment was 3.7+/-1.5 mm. A 6F long sheath was used for delivery of the ADO. Follow-up evaluation was performed with color flow mapping of the main pulmonary artery within 24 h and at 1 and 3 months after closure.
RESULTS: Twenty three of the 24 patients had successful device placement. Angiography showed that 7 patients had complete immediate closure, 14 had a trace shunt (foaming through the device with no jet), and 2 had a small residual shunt (with a jet). Within 24 h, color Doppler revealed complete closure in all patients. The unsuccessful attempt was during an initial trial with a prototype that has been modified. The median fluoroscopy time was 13.5 min (range 6.3 to 47). All patients were discharged home the next day. There were no complications. Of the 23 patients, 21 completed the 1-month follow-up, all (95% confidence interval [CI] 86% to 100%) with complete closure, and 18 of 23 patients completed the 3-month follow-up, also all (95% CI 83% to 100%) with complete closure.
CONCLUSIONS: Anterograde transcatheter closure using the new ADO is an effective therapy for patients with a PDA diameter up to 6 mm. Further clinical trials are underway.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9525563     DOI: 10.1016/s0735-1097(98)00013-8

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  60 in total

1.  Interventional pediatric cardiology: device closures.

Authors:  J L Wilkinson
Journal:  Indian J Pediatr       Date:  2000-07       Impact factor: 1.967

Review 2.  Interventional paediatric cardiology.

Authors:  Kevin P Walsh
Journal:  BMJ       Date:  2003-08-16

3.  Transcatheter occlusion of a large left coronary artery to right superior vena cava fistula using the Amplatzer duct occluder device.

Authors:  J Al-Ata; M Amin; M O Galal; A Kouatli; A Hussain
Journal:  Pediatr Cardiol       Date:  2003-10-31       Impact factor: 1.655

4.  Percutanous closure of patent ductus arteriosus in small infants of less than 8 kg body weight using different devices.

Authors:  Walter Knirsch; Nikolaus A Haas; Martin A G Lewin; Ingo Dähnert; Deniz Kececioglu; Felix Berger; Frank Uhlemann
Journal:  Eur J Pediatr       Date:  2004-10       Impact factor: 3.183

Review 5.  Closing down: transcatheter closure of intracardiac defects and vessel embolisations.

Authors:  Jean-François Piéchaud
Journal:  Heart       Date:  2004-12       Impact factor: 5.994

6.  Transcatheter occlusion of the patent ductus arteriosus: a comparison of two devices.

Authors:  M K El Mallah; A J Sands; F A Casey; B G Craig; H C Mulholland
Journal:  Ir J Med Sci       Date:  2002 Jul-Sep       Impact factor: 1.568

7.  Transcatheter closure of patent ductus arteriosus and interruption of inferior vena cava with azygous continuation using an Amplatzer duct occluder.

Authors:  Sadiq Al-Hamash
Journal:  Pediatr Cardiol       Date:  2006-08-23       Impact factor: 1.655

8.  Transcatheter closure of a large patent ductus arteriosus in a young child using the Amplatzer duct occluder.

Authors:  S-L Jan; B Hwang; Y-C Fu; C-S Chi
Journal:  Pediatr Cardiol       Date:  2005 Sep-Oct       Impact factor: 1.655

9.  Transcatheter closure with use of the SHSMA occluder in 180 patients with congenital heart defects: preliminary results.

Authors:  Wan-Feng Sun; Zhi-Feng Dong; Kaizheng Gong; Guo-Pei Zhang; Ting Cui; Yu-Dong Xia; Jing Dong; Yuan Shen
Journal:  Tex Heart Inst J       Date:  2010

Review 10.  Use of Amplatzer Vascular Plug (AVP) in emergency embolisation: preliminary experience and review of literature.

Authors:  Monica Mangini; Domenico Laganà; Federico Fontana; Andrea Ianniello; Paolo Nicotera; Maria Petullà; Gianpaolo Carrafiello; Carlo Fugazzola
Journal:  Emerg Radiol       Date:  2008-01-10
View more