Multicenter evaluation of the Paragon CZE 2000 capillary zone electrophoresis system for serum protein electrophoresis and monoclonal component typing.

J Bienvenu1, M S Graziani, F Arpin, H Bernon, C Blessum, C Marchetti, G Righetti, M Somenzini, G Verga, F Aguzzi.   

Abstract

Serum protein electrophoresis and typing of monoclonal components (MCs) are routine but time-consuming and technically demanding assays. We evaluated capillary electrophoresis (Paragon CZE 2000) for automation of the two assays. CZE and cellulose acetate electrophoresis gave similar data on 794 samples. Within-run and between-run CVs were < 2% for albumin and gamma-globulins and 4-7% for alpha 1-, alpha 2-, and beta-globulins. Bilirubin, hemoglobin, triglycerides, and fibrinogen were found not to interfere. No carryover by capillaries was detected. The detection limit for MC was < 0.5 g/L. MC assessment by immunosubtraction on 403 samples identified the monoclonal type in all samples with peak concentrations > 10 g/L; only 50% of MCs that could not be quantified by densitometric scan were typed.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9510868

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  3 in total

Review 1.  Clinical applications of capillary electrophoresis. Status at the new millennium.

Authors:  M A Jenkins
Journal:  Mol Biotechnol       Date:  2000-07       Impact factor: 2.695

2.  Homogeneous immunosubtraction integrated with sample preparation enabled by a microfluidic format.

Authors:  Akwasi A Apori; Amy E Herr
Journal:  Anal Chem       Date:  2011-03-04       Impact factor: 6.986

3.  Measurement of serum albumin by capillary zone electrophoresis, bromocresol green, bromocresol purple, and immunoassay methods.

Authors:  E B Duly; S Grimason; P Grimason; G Barnes; T R Trinick
Journal:  J Clin Pathol       Date:  2003-10       Impact factor: 3.411

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.