Rotary bending fatigue of coils and wires used in cardiac lead design.

P A Altman1, J M Meagher, D W Walsh, D A Hoffmann.   

Abstract

Implantable electrical conductor wires and coils of MP35N and drawn MP35N tube filled with a silver core (DFT) were tested in rotary bending to evaluate fatigue characteristics and failure mechanisms. Detailed stress analysis enabled evaluation of helical coil fatigue in bending with respect to theoretical stress. The relations between the maximum cyclic stress (S) and the cycles to failure (N) were obtained by introducing cyclic stresses on over 600 samples. Microscopic evaluation using scanning electron microscopy was performed to identify failure mechanisms. The results show that the stress analysis presented is a viable means for assessing the stress state in a composite coil structure in rotary loop testing, comparable S-N fatigue data sets for MP35N and DFT conductors are not consistently significantly different, S-N relationships for coils and wires of the MP35N and DFT appear to lie on the same curves, and failures of all sample types initiate on the outside surface of the wire. Differences in crack propagation and final fracture are shown as a function of sample type and cyclic stress.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9509341     DOI: 10.1002/(sici)1097-4636(199821)43:1<21::aid-jbm3>3.0.co;2-o

Source DB:  PubMed          Journal:  J Biomed Mater Res        ISSN: 0021-9304


  5 in total

1.  Comparing Rotary Bend Wire Fatigue Test Methods at Different Test Speeds.

Authors:  Jason D Weaver; Erick J Gutierrez
Journal:  J Mater Eng Perform       Date:  2015-12       Impact factor: 1.819

2.  Analytical Modeling for Computing Lead Stress in a Novel Epicardial Micropacemaker.

Authors:  Li Zhou; Yaniv Bar-Cohen; Raymond A Peck; Giorgio V Chirikian; Brett Harwin; Ramen H Chmait; Jay D Pruetz; Michael J Silka; Gerald E Loeb
Journal:  Cardiovasc Eng Technol       Date:  2017-01-09       Impact factor: 2.495

3.  The development of neural stimulators: a review of preclinical safety and efficacy studies.

Authors:  Robert K Shepherd; Joel Villalobos; Owen Burns; David A X Nayagam
Journal:  J Neural Eng       Date:  2018-05-14       Impact factor: 5.379

4.  Tension and fatigue behavior of 316LVM 1x7 multi-strand cables used as implantable electrodes.

Authors:  John J Lewandowski; Ravikumar Varadarajan; Brian Smith; Chris Tuma; Mostafa Shazly; Luciano O Vatamanu
Journal:  Mater Sci Eng A Struct Mater       Date:  2008-07-15       Impact factor: 5.234

5.  Three-Dimensional Analysis of the In Vivo Motion of Implantable Cardioverter Defibrillator Leads.

Authors:  Tamas Szili-Torok; Jens Rump; Torsten Luther; Sing-Chien Yap
Journal:  Cardiovasc Eng Technol       Date:  2021-06-29       Impact factor: 2.495

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.