Literature DB >> 9507547

Comparison of right- and left-sided pectoral implantation parameters with the Jewel active can cardiodefibrillator. The World Wide Jewel Investigators.

G C Flaker1, R Tummala, J Wilson.   

Abstract

A total of 1,207 patients received a Medtronic Jewel active can ICD (models 7218C, 7219C), with a Transvene lead in 97 centers in Europe and North America. Nineteen implants were from the right pectoral region. Patients with right-sided ICDs did not differ in terms of mean age, % male, left ventricular ejection fraction, New York Heart Association Functional Class, antiarrhythmic drug therapy, indication for the implantable cardioverter defibrillator, and R wave values at implantation, but tended to have slightly higher pacing thresholds (1.2 +/- 0.5 V vs 1.0 +/- 0.6 V, P = 0.012) and higher defibrillation thresholds (14.7 +/- 6.4 J vs 11.5 +/- 6 J, P = 0.11) compared with patients with left sided implants. Patients with right-sided implants had a longer implantation time compared with patients with left-sided implants (118 +/- 70 minutes vs 91 +/- 46 minutes, P = 0.074). In follow-up, 5 patients with right-sided implantation received successful therapy for either ventricular fibrillation, (8 episodes) or ventricular tachycardia (5 episodes). No ineffective therapy from the device was delivered in any patients with right-sided implantation. Right-sided pectoral implants are feasible with the Medtronic Jewel active can ICD.

Entities:  

Mesh:

Year:  1998        PMID: 9507547     DOI: 10.1111/j.1540-8159.1998.tb00070.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  2 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

2.  Increased defibrillation threshold with right-sided active pectoral can.

Authors:  P R Roberts; S Allen; T Betts; J M Morgan; J F Urban; T Whitman; D E Euler; M J Kallok
Journal:  J Interv Card Electrophysiol       Date:  2000-04       Impact factor: 1.900

  2 in total