Literature DB >> 9507542

ICD restudy: results and potential benefit from routine predischarge and 2-month evaluation.

S L Higgins1, D H Rich, J R Haygood, J Barone, S L Greer, D B Meyer.   

Abstract

Evaluation of ICD function can now be performed noninvasively with intravenous sedation. To determine the value of follow-up electrophysiological studies for ICD implants, we performed a retrospective review of predischarge and 2-month ICD re studies, identifying critical problems uncovered. Of the 123 patients implanted, 122 had a predischarge study, 105 had both predischarge and elective 2-month follow-up studies, and 1 patient expired prior to restudy. Patients who underwent 2-month studies for nonelective indications (e.g., frequent shocks) were excluded from analysis. Programming changes were made in 62% of the predischarge studies (n = 122) and 70% of the elective 2-month studies (n = 105). The average number of programming changes per study was 1.3 for predischarge testing and 1.1 for 2-month testing. The most common changes at predischarge testing were adjustment of the tachyarrhythmia rate cutoff (35%) and at 2-month study, reprogramming of bradycardia pacing parameters (41%). Of the patients who underwent both predischarge and 2-month testing, 91% had programming changes in at least one of their re studies. Of 227 re studies performed, 18 studies in 14 patients yielded 24 critical findings which included: DFT increases to > or = 25 J (n = 13); sensing abnormalities of induced ventricular arrhythmia (n = 6); dislodged lead (n = 2); and serious pacemaker interactions (n = 3). Six of these critical cases (5% of total patients) required reoperation. The data suggests that routine ICD restudy is a valuable tool for management of the ICD patient. Additionally, ICD restudy is likely to increase the diagnostic yield of clinically silent critical system problems that could result in device failure.

Entities:  

Mesh:

Year:  1998        PMID: 9507542     DOI: 10.1111/j.1540-8159.1998.tb00065.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  3 in total

1.  Indications for predismissal testing with arrhythmia-induction in patients receiving an implantable cardioverter defibrillator.

Authors:  G Christ; R Becker; F Voss; K Kelemen; J Senges-Becker; M Hauck; W Schoels; I Bald; H A Katus; A Bauer
Journal:  Clin Res Cardiol       Date:  2007-06-27       Impact factor: 5.460

2.  Clinical implications of left superior vena cava persistence in candidates for pacemaker or cardioverter-defibrillator implantation.

Authors:  Mauro Biffi; Matteo Bertini; Matteo Ziacchi; Cristian Martignani; Cinzia Valzania; Igor Diemberger; Angelo Branzi; Giuseppe Boriani
Journal:  Heart Vessels       Date:  2009-04-01       Impact factor: 2.037

3.  Defibrillation testing of the implantable cardioverter defibrillator: when, how, and by whom?

Authors:  Luis A Pires
Journal:  Indian Pacing Electrophysiol J       Date:  2007-08-01
  3 in total