Literature DB >> 9486547

Reproducibility of optic disk topographic measurements with the Topcon ImageNet and the Heidelberg Retina Tomograph.

A Azuara-Blanco1, A Harris, L B Cantor.   

Abstract

OBJECTIVE: To compare the reproducibility of optic disk measurements provided by an image analyzer and a scanning laser tomograph.
METHODS: Ten images of the same eye of 10 normal volunteers were taken with the Heidelberg Retina Tomograph and with the Topcon ImageNet. Intraclass correlation coefficient (ICC) and coefficient of variation (CV) were used to evaluate the reproducibility of the measurements.
RESULTS: Eleven parameters were analyzed with the Topcon ImageNet. Six parameters (55%) had ICC greater than 90%. Four parameters (36%) had CV less than 10%. Twelve parameters were evaluated with the Heidelberg Retina Tomograph. Nine parameters (75%) had ICC over 90%. Nine parameters (75%) had CV less than 10%.
CONCLUSION: Both systems provided reproducible data. The optic disk parameters provided by the Heidelberg Retina Tomograph had a better reproducibility than those obtained from the Topcon ImageNet.

Mesh:

Year:  1998        PMID: 9486547     DOI: 10.1159/000027286

Source DB:  PubMed          Journal:  Ophthalmologica        ISSN: 0030-3755            Impact factor:   3.250


  5 in total

1.  Comparison between laser scanning tomography and computerised image analysis of the optic disc.

Authors:  A Azuara-Blanco; G L Spaeth; J Nicholl; I M Lanzl; J J Augsburger
Journal:  Br J Ophthalmol       Date:  1999-03       Impact factor: 4.638

2.  The Humphrey optical coherence tomography scanner: quantitative analysis and reproducibility study of the normal human retinal nerve fibre layer.

Authors:  A L Jones; N J Sheen; R V North; J E Morgan
Journal:  Br J Ophthalmol       Date:  2001-06       Impact factor: 4.638

Review 3.  Test-retest variability in structural parameters measured with glaucoma imaging devices.

Authors:  Makoto Araie
Journal:  Jpn J Ophthalmol       Date:  2012-11-09       Impact factor: 2.447

4.  Optic disc topographic parameters measured in the normal cynomolgus monkey by confocal scanning laser tomography.

Authors:  T Taniguchi; M Shimazawa; M Araie; G Tomita; M Sasaoka; Y Kitazawa; H Hara
Journal:  Br J Ophthalmol       Date:  2005-08       Impact factor: 4.638

5.  A genome-wide association study of optic disc parameters.

Authors:  Wishal D Ramdas; Leonieke M E van Koolwijk; M Kamran Ikram; Nomdo M Jansonius; Paulus T V M de Jong; Arthur A B Bergen; Aaron Isaacs; Najaf Amin; Yurii S Aulchenko; Roger C W Wolfs; Albert Hofman; Fernando Rivadeneira; Ben A Oostra; Andre G Uitterlinden; Pirro Hysi; Christopher J Hammond; Hans G Lemij; Johannes R Vingerling; Caroline C W Klaver; Cornelia M van Duijn
Journal:  PLoS Genet       Date:  2010-06-10       Impact factor: 5.917

  5 in total