Literature DB >> 9483443

A 433 MHz Lucite cone waveguide applicator for superficial hyperthermia.

G C van Rhoon1, P J Rietveld, J van der Zee.   

Abstract

The effective field size (EFS, SAR > or = 50% of the maximum SAR at 1 cm depth) of a conventional 433 MHz water filled waveguide applicator (32 cm2, aperture area 100 cm2) has been increased by: (1) replacement of the two diverging brass side walls which are parallel to the direction of the electric field by Lucite walls; and (2) Placement of a heterogeneous permittivity in the centre of the aperture. SAR distributions were measured at several depths in layered fat-muscle phantoms. With Lucite side walls the SAR distribution becomes wider in the H-plane of the aperture, resulting in a circular SAR distribution. In this situation the EFS is 67 cm2. Additional insertion of a PVC cone with a top angle of 15 degrees at the centre of the aperture increases the EFS to 91 +/- 6 cm2 for a waterbolus of 18 x 18 x 1 cm3. The experiments also demonstrated that the resulting EFS is affected by the waterbolus size and shape. Calorimetric measurements showed that the efficiency of the improved applicator is comparable to the efficiency of the conventional water filled waveguide applicator, 50 and 56% respectively. The modifications reported provide a simple and inexpensive means to increase the EFS and can be easily implemented in water filled waveguide applicators.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9483443     DOI: 10.3109/02656739809018211

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  10 in total

1.  Performance evaluation of a conformal thermal monitoring sheet sensor array for measurement of surface temperature distributions during superficial hyperthermia treatments.

Authors:  K Arunachalam; P Maccarini; T Juang; C Gaeta; P R Stauffer
Journal:  Int J Hyperthermia       Date:  2008-06       Impact factor: 3.914

2.  Procedure for creating a three-dimensional (3D) model for superficial hyperthermia treatment planning.

Authors:  Marianne Linthorst; Tomas Drizdal; Hans Joosten; Gerard C van Rhoon; Jacoba van der Zee
Journal:  Strahlenther Onkol       Date:  2011-11-25       Impact factor: 3.621

Review 3.  Heating technology for malignant tumors: a review.

Authors:  H Petra Kok; Erik N K Cressman; Wim Ceelen; Christopher L Brace; Robert Ivkov; Holger Grüll; Gail Ter Haar; Peter Wust; Johannes Crezee
Journal:  Int J Hyperthermia       Date:  2020       Impact factor: 3.914

4.  Vesicoureteral reflux in children: a phantom study of microwave heating and radiometric thermometry of pediatric bladder.

Authors:  Yngve Birkelund; Øystein Klemetsen; Svein K Jacobsen; Kavitha Arunachalam; Paolo Maccarini; Paul R Stauffer
Journal:  IEEE Trans Biomed Eng       Date:  2011-09-06       Impact factor: 4.538

5.  Evaluation of CEM43 degrees CT90 thermal dose in superficial hyperthermia: a retrospective analysis.

Authors:  Maarten de Bruijne; Bronno van der Holt; Gerard C van Rhoon; Jacoba van der Zee
Journal:  Strahlenther Onkol       Date:  2010-07-29       Impact factor: 3.621

6.  The Kadota Fund International Forum 2004--clinical group consensus.

Authors:  J van der Zee; Z Vujaskovic; M Kondo; T Sugahara
Journal:  Int J Hyperthermia       Date:  2008-03       Impact factor: 3.914

7.  Flow patterns and heat convection in a rectangular water bolus for use in superficial hyperthermia.

Authors:  Yngve Birkelund; Svein Jacobsen; Kavitha Arunachalam; Paolo Maccarini; Paul R Stauffer
Journal:  Phys Med Biol       Date:  2009-06-03       Impact factor: 3.609

8.  Microwave applicator for hyperthermia treatment on in vivo melanoma model.

Authors:  Paolo Togni; Jan Vrba; Luca Vannucci
Journal:  Med Biol Eng Comput       Date:  2009-12-24       Impact factor: 2.602

Review 9.  Quality assurance guidelines for superficial hyperthermia clinical trials : II. Technical requirements for heating devices.

Authors:  Hana Dobšíček Trefná; Johannes Crezee; Manfred Schmidt; Dietmar Marder; Ulf Lamprecht; Michael Ehmann; Jacek Nadobny; Josefin Hartmann; Nicolleta Lomax; Sultan Abdel-Rahman; Sergio Curto; Akke Bakker; Mark D Hurwitz; Chris J Diederich; Paul R Stauffer; Gerard C Van Rhoon
Journal:  Strahlenther Onkol       Date:  2017-03-01       Impact factor: 3.621

10.  Modelling Curved Contact Flexible Microstrip Applicators for Patient-Specific Superficial Hyperthermia Treatment Planning.

Authors:  H Petra Kok; Jort Groen; Akke Bakker; Johannes Crezee
Journal:  Cancers (Basel)       Date:  2020-03-11       Impact factor: 6.639