Effect of registration errors between transmission and emission scans on a SPECT system using sequential scanning.

C D Stone1, J W McCormick, D R Gilland, K L Greer, R E Coleman, R J Jaszczak.   

Abstract

UNLABELLED: The purpose of this study was to evaluate the effects of patient motion on nonuniform attenuation correction of cardiac SPECT when the transmission and emission scans were performed sequentially. By using a sequential protocol rather than doing the scans simultaneously, contamination from the emission scan into the transmission scan could be eliminated, but registration of the two scans become a concern.
METHODS: Transmission and emission scans were acquired using both an anthropomorphic thorax phantom containing a cardiac insert and a human volunteer. The types of motion considered were transverse shifts, axial shifts and rotations that occur in the time period between the transmission and emission scan. For this study, the various types of motion were simulated in the data. Both the transmission and emission data were reconstructed using filtered backprojection. A single-iteration Chang algorithm, modified for nonuniform attenuation correction, was used to further process the emission data. To evaluate the effects of motion errors, circumferential profiles, all normalized to the same scale, were generated for each case. The cardiac images reconstructed using registered data were considered references. Error profiles were generated by subtracting misaligned images from the reference and then normalizing the difference by the reference. For comparison purposes, an error profile was generated for the case in which no attenuation correction was performed.
RESULTS: It was found that, for transverse and axial shifts of 2.9 cm, the normalized myocardial SPECT activity was decreased in certain regions of the heart by 20%-35%. For a 12 degrees rotational shift, the error was on the order of 10%-20%, compared to a normalized variation of 20%-25% in the image with no attenuation correction.
CONCLUSION: The results indicate that registration errors of 2-3 cm can seriously affect image quality in both the phantom and human images.

Entities:  

Mesh:

Year:  1998        PMID: 9476952

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  8 in total

Review 1.  Effect of motion on cardiac SPECT imaging: recognition and motion correction.

Authors:  J Fitzgerald; P G Danias
Journal:  J Nucl Cardiol       Date:  2001 Nov-Dec       Impact factor: 5.952

2.  Assessment of geometrical distortion and activity distribution after attenuation correction: A SPECT phantom study.

Authors:  Sara Lappi; Simonetta Lazzari; Graziella Sarti; Pierluigi Pieri
Journal:  J Nucl Cardiol       Date:  2002 Sep-Oct       Impact factor: 5.952

3.  Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT.

Authors:  Sibyll Goetze; Richard L Wahl
Journal:  J Nucl Cardiol       Date:  2007-04       Impact factor: 5.952

4.  Use of MRI to assess the prediction of heart motion with gross body motion in myocardial perfusion imaging by stereotracking of markers on the body surface.

Authors:  Michael A King; Joyoni Dey; Karen Johnson; Paul Dasari; Joyeeta M Mukherjee; Joseph E McNamara; Arda Konik; Cliff Lindsay; Shaokuan Zheng; Dennis Coughlin
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

5.  Is SPECT or CT Based Attenuation Correction More Quantitatively Accurate for Dedicated Breast SPECT Acquired with Non-Traditional Trajectories?

Authors:  Kristy L Perez; Steve D Mann; Jan H Pachon; Priti Madhav; Martin P Tornai
Journal:  IEEE Nucl Sci Symp Conf Rec (1997)       Date:  2010 Oct-Nov

Review 6.  Accuracy and precision of radioactivity quantification in nuclear medicine images.

Authors:  Eric C Frey; John L Humm; Michael Ljungberg
Journal:  Semin Nucl Med       Date:  2012-05       Impact factor: 4.446

7.  Fisher information analysis of list-mode SPECT emission data for joint estimation of activity and attenuation distribution.

Authors:  Ashequr Rahman; Yansong Zhu; Eric Clarkson; Matthew A Kupinski; Eric C Frey; Abhinav K Jha
Journal:  Inverse Probl       Date:  2020-08-20       Impact factor: 2.407

Review 8.  An Overview on Image Registration Techniques for Cardiac Diagnosis and Treatment.

Authors:  Azira Khalil; Siew-Cheok Ng; Yih Miin Liew; Khin Wee Lai
Journal:  Cardiol Res Pract       Date:  2018-08-08       Impact factor: 1.866

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.