Literature DB >> 9473160

Evaluation of a new portable system for cord lactate determination.

J Yam1, S Chua, K Razvi, S Arulkumaran.   

Abstract

To evaluate a new reflectance photometer (Accusport) for blood lactate determination, a prospective non-interventional study was performed on blood obtained from the umbilical cord artery immediately after delivery of the baby in 109 consecutive women. For each sample, the results of cord arterial blood lactate values measured with the reflectance photometer were compared with those obtained using a reference method in the laboratory. The values obtained using Accusport were comparable in accuracy with those obtained using a laboratory reference method (Ektachem 9501RC). The instrument provides a simple and reliable method for determining cord arterial blood lactate levels in a small blood sample.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9473160     DOI: 10.1159/000009919

Source DB:  PubMed          Journal:  Gynecol Obstet Invest        ISSN: 0378-7346            Impact factor:   2.031


  1 in total

1.  Portable lactate analyzer for measuring lactate in cerebrospinal fluid (CSF) and plasma - method-comparison evaluations.

Authors:  Sérgio Monteiro de Almeida; Jennifer Marquie-Beck; Archana Bhatt; Scott Letendre; Allen McCutchan; Ron Ellis
Journal:  Arq Neuropsiquiatr       Date:  2014-07       Impact factor: 1.420