Literature DB >> 9472833

Diagnostic x-ray spectra: a comparison of spectra generated by different computational methods with a measured spectrum.

M Bhat1, J Pattison, G Bibbo, M Caon.   

Abstract

A number of computer codes, developed using semi-empirical models, are available to compute x-ray spectra from a tungsten target for different tube parameters. In this study x-ray spectra measured with a high-purity germanium detector are compared with those computed using the empirical models and previously published measured data. The computer codes used to generate the spectra are based on models proposed by Birch et al. and Tucker et al. The measured x-ray spectra agreed well with the computed x-ray spectra using the model of Tucker et al. whereas the model of Birch et al. produced a "harder" x-ray spectrum compared to the measured spectra. Our measured x-ray spectra compared well with the previously published measured spectral data of Fewell et al.

Mesh:

Substances:

Year:  1998        PMID: 9472833     DOI: 10.1118/1.598170

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  7 in total

1.  A grating-based single-shot x-ray phase contrast and diffraction method for in vivo imaging.

Authors:  Eric E Bennett; Rael Kopace; Ashley F Stein; Han Wen
Journal:  Med Phys       Date:  2010-11       Impact factor: 4.071

2.  Analytical reconstructions of intensity modulated x-ray phase-contrast imaging of human scale phantoms.

Authors:  Bartłomiej Włodarczyk; Jakub Pietrzak
Journal:  Biomed Opt Express       Date:  2015-10-05       Impact factor: 3.732

3.  Fourier X-ray scattering radiography yields bone structural information.

Authors:  Han Wen; Eric E Bennett; Monica M Hegedus; Stanislas Rapacchi
Journal:  Radiology       Date:  2009-04-29       Impact factor: 11.105

4.  Use of a laser-guided collimation system to perform direct kilovoltage x-ray spectra measurements on a linear accelerator onboard imager.

Authors:  Zachary Grelewicz; Andrew H Belcher; Rodney D Wiersma
Journal:  Med Phys       Date:  2018-10-09       Impact factor: 4.071

5.  CT scanner x-ray spectrum estimation from transmission measurements.

Authors:  Xinhui Duan; Jia Wang; Lifeng Yu; Shuai Leng; Cynthia H McCollough
Journal:  Med Phys       Date:  2011-02       Impact factor: 4.071

6.  Real-time dosimeter employed to evaluate the half-value layer in CT.

Authors:  Sarah E McKenney; J Anthony Seibert; George W Burkett; Dale Gelskey; Paul B Sunde; James D Newman; John M Boone
Journal:  Phys Med Biol       Date:  2013-12-19       Impact factor: 3.609

7.  Semiempirical, parameterized spectrum estimation for x-ray computed tomography.

Authors:  Paul FitzGerald; Stephen Araujo; Mingye Wu; Bruno De Man
Journal:  Med Phys       Date:  2021-03-16       Impact factor: 4.071

  7 in total