Literature DB >> 9457208

Changes in high-dose-rate tandem and ovoid applicator positions during treatment in an unfixed brachytherapy system.

H T Pham1, Y Chen, E Rouby, R A Lustig, P E Wallner.   

Abstract

PURPOSE: To measure the change in applicator position during treatment in an unfixed high-dose-rate (HDR) brachytherapy system and to evaluate the effect of the shifts on dose calculations.
MATERIALS AND METHODS: Posttreatment localization radiographs were obtained for 47 HDR treatments (26 tandem and ovoid applicators, 21 ovoids-only applicators). The authors measured the change in applicator position relative to the patient's bone anatomy. Doses to the target and critical structures were calculated for posttreatment applicator positions for comparison.
RESULTS: Average displacements of the tandem and ovoid applicators in anteroposterior dimension were 5 and 4 mm for the tandem and ovoids, respectively. Anterior displacement occurred twice as often as posterior displacement. The average lateral and longitudinal shifts were less. Less displacement was observed with ovoids-only insertions. The largest displacement for ovoids-only applicators was 3 mm in the anteroposterior dimension. A high bladder dose difference (17.4%) for tandem and ovoid applications correlated with anterior shifts of the applicators.
CONCLUSION: Patient movement in an unfixed HDR brachytherapy system can displace the applicators, especially the tandem. Anterior shifts correlate with high bladder dose differences. Immobilization of the patient's hips and legs, as well as stabilization of applicators, would reduce these shifts.

Entities:  

Mesh:

Year:  1998        PMID: 9457208     DOI: 10.1148/radiology.206.2.9457208

Source DB:  PubMed          Journal:  Radiology        ISSN: 0033-8419            Impact factor:   11.105


  6 in total

1.  Radiation protection in radiological imaging: a survey of imaging modalities used in Japanese institutions for verifying applicator placements in high-dose-rate brachytherapy.

Authors:  Hiroyuki Okamoto; Satoshi Kito; Naoki Tohyama; Shunsuke Yonai; Ryu Kawamorita; Masaru Nakamura; Takahiro Fujimoto; Syoji Tani; Akihiro Yomoda; Toru Isobe; Hiroshi Furukawa; Kikuo Kotaka; Jun Itami; Hitoshi Ikushima; Takushi Dokiya; Yoshiyuki Shioyama
Journal:  J Radiat Res       Date:  2021-01-01       Impact factor: 2.724

2.  Positional variation of applicators during low dose rate intracavitary brachytherapy for cervical cancer: a prospective study.

Authors:  Kirthi Koushik; Ramesh Bilimagga; Nirmala Rao; M G Janaki; Arul Ponni; A G Rajeev
Journal:  J Contemp Brachytherapy       Date:  2010-10-13

3.  Evaluation of variation in dose of organs at risk in intracavitary brachytherapy of cervical cancer - a prospective study.

Authors:  Brahmacharimayum Arunkumar Sharma; Th Tomcha Singh; L Jaichand Singh
Journal:  J Contemp Brachytherapy       Date:  2011-03-31

4.  Commissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets.

Authors:  Ryan McMahon; Tingliang Zhuang; Beverly A Steffey; Haijun Song; Oana I Craciunescu
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

5.  Magnitude and Implications of Interfraction Variations in Organ Doses during High Dose Rate Brachytherapy of Cervix Cancer: A CT Based Planning Study.

Authors:  Santam Chakraborty; Firuza D Patel; Vijay M Patil; Arun S Oinam; Suresh C Sharma
Journal:  ISRN Oncol       Date:  2014-02-03

6.  Reduction of applicator displacement in MR/CT-guided cervical cancer HDR brachytherapy by the use of patient hover transport system.

Authors:  Megan Andrew; Yusung Kim; Timothy Ginader; Brian J Smith; Wenqing Sun; Dongxu Wang
Journal:  J Contemp Brachytherapy       Date:  2018-02-26
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.