Accuracy of a cerebral oximeter in healthy volunteers under conditions of isocapnic hypoxia.

L C Henson1, C Calalang, J A Temp, D S Ward.   

Abstract

BACKGROUND: A cerebral oximeter measures oxygen saturation of brain tissue noninvasively by near infrared spectroscopy. The accuracy of a commercially available oximeter was tested in healthy volunteers by precisely controlling end-tidal oxygen (P[ET]O2) and carbon dioxide (P[ET]CO2) tensions to alter global cerebral oxygen saturation.
METHODS: In 30 healthy volunteers, dynamic end-tidal forcing was used to produce step changes in P[ET]O2 resulting in arterial saturation ranging from approximately 70% to 100% under conditions of controlled normocapnia (each person's resting P[ET]CO2) or hypercapnia (resting plus 7-10 mmHg). Blood arterial (SaO2) and jugular bulb venous (S[jv]O2) saturations during each P(ET)O2 interval were determined by co-oximetry. The cerebral oximeter reading (rSO2) and an estimated jugular venous saturation (S[jv]O2), derived from a combination of SaO2 and rSO2, were compared with the measured S(jv)O2.
RESULTS: The S(jv)O2 was significantly higher with hypercapnia than with normocapnia for the same SaO2. The rSO2 and S(jv)O2 were both highly correlated with S(jv)O2 for individual volunteers (mean r2 = 0.91 for each relation); however, the slopes and intercepts varied widely among volunteers. In three of them, the cerebral oximeter substantially underestimated the measured S(jv)O2.
CONCLUSIONS: During isocapnic hypoxia in healthy persons, cerebral oxygenation as estimated by near infrared spectroscopy precisely tracks changes in measured S(jv)O2 within individuals, but the relation exhibits a wide range of slopes and intercepts. Therefore the clinical utility of the device is limited to situations in which tracking trends in cerebral oxygenation would be acceptable.

Entities:  

Mesh:

Substances:

Year:  1998        PMID: 9447856     DOI: 10.1097/00000542-199801000-00011

Source DB:  PubMed          Journal:  Anesthesiology        ISSN: 0003-3022            Impact factor:   7.892


  19 in total

1.  Transabdominal near infrared oximetry of hypoxic stress in fetal sheep brain in utero.

Authors:  Regine Choe; Turgut Durduran; Guoqiang Yu; Mark J M Nijland; Britton Chance; Arjun G Yodh; Nirmala Ramanujam
Journal:  Proc Natl Acad Sci U S A       Date:  2003-10-16       Impact factor: 11.205

2.  Cerebral vasoreactivity during hypercapnia is reset by augmented sympathetic influence.

Authors:  Peizhen Zhang; Guoyuan Huang; Xiangrong Shi
Journal:  J Appl Physiol (1985)       Date:  2010-11-11

Review 3.  The clinical validity of the absolute value of near infrared spectroscopy.

Authors:  Kenji Yoshitani; Yoshihiko Ohnishi
Journal:  J Anesth       Date:  2008-11-15       Impact factor: 2.078

4.  Suppression of cerebral hemodynamics is associated with reduced functional capacity in patients with heart failure.

Authors:  Tieh-Cheng Fu; Chao-Hung Wang; Chih-Chin Hsu; Wen-Jin Cherng; Shu-Chun Huang; Jong-Shyan Wang
Journal:  Am J Physiol Heart Circ Physiol       Date:  2011-01-28       Impact factor: 4.733

Review 5.  In vivo validation of cerebral near-infrared spectroscopy: a review.

Authors:  Amalie la Cour; Gorm Greisen; Simon Hyttel-Sorensen
Journal:  Neurophotonics       Date:  2018-11-27       Impact factor: 3.593

6.  Effect of anesthetic methods on cerebral oxygen saturation in elderly surgical patients: prospective, randomized, observational study.

Authors:  Aerina Lee; Sung-Hoon Kim; Jeong-Yeon Hong; Jai-Hyun Hwang
Journal:  World J Surg       Date:  2012-10       Impact factor: 3.352

7.  The accuracy of a near-infrared spectroscopy cerebral oximetry device and its potential value for estimating jugular venous oxygen saturation.

Authors:  Keita Ikeda; David B MacLeod; Hilary P Grocott; Eugene W Moretti; Warwick Ames; Charles Vacchiano
Journal:  Anesth Analg       Date:  2014-12       Impact factor: 5.108

8.  Noninvasive cerebral oximeter as a surrogate for mixed venous saturation in children.

Authors:  Adnan T Bhutta; Jesse W Ford; James G Parker; Parthak Prodhan; Eudice E Fontenot; Paul M Seib; Brittany I Stroope; Elizabeth A Frazier; Michael L Schmitz; Jonathan J Drummond-Webb; William R Morrow
Journal:  Pediatr Cardiol       Date:  2007-01-09       Impact factor: 1.655

9.  Estimation of jugular venous O2 saturation from cerebral oximetry or arterial O2 saturation during isocapnic hypoxia.

Authors:  M B Kim; D S Ward; C R Cartwright; J Kolano; S Chlebowski; L C Henson
Journal:  J Clin Monit Comput       Date:  2000       Impact factor: 2.502

10.  The change in regional cerebral oxygen saturation after stellate ganglion block.

Authors:  Hyeon Min Park; Tae Wan Kim; Hong Gyu Choi; Kyung Bong Yoon; Duck Mi Yoon
Journal:  Korean J Pain       Date:  2010-05-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.