A new Diamentor for measuring kerma-area product and air-kerma simultaneously.

H Gfirtner1, F E Stieve, J Wild.   

Abstract

The new M4KDK Diamentor (PTW-Freiburg) for measuring the air-kerma-area product (KAP) and air kerma (AK) simultaneously was evaluated. The special design of the Diamentor chamber influences the variation of the calibration factor with respect to the field size somewhat differently to what is the case with the classic Diamentor. It is therefore recommended that calibration be performed for a larger field size. If the M4KDK Diamentor is calibrated for a field size of 3.8 cm x 3.8 cm (at the chamber), an uncertainty in the AK value of better than 15% for a field size larger than the calibration field, can be expected. The influence of backscattered radiation on the calibration factor is equivalent to that for the classic Diamentor chamber. The variation in response in accordance with tube potential, is also comparable with that of the classic Diamentor. Dramatic changes in response have to be expected if additional filters are introduced in close proximity to the Diamentor chamber.

Mesh:

Year:  1997        PMID: 9434978     DOI: 10.1118/1.598109

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071