Respiratory comfort of automatic tube compensation and inspiratory pressure support in conscious humans.

J Guttmann1, H Bernhard, G Mols, A Benzing, P Hofmann, C Haberthür, D Zappe, B Fabry, K Geiger.   

Abstract

OBJECTIVE: To compare the new mode of ventilatory support, which we call automatic tube compensation (ATC), with inspiratory pressure support (IPS) with respect to perception of respiratory comfort. ATC unloads the resistance of the endotracheal tube (ETT) in inspiration by increasing the airway pressure, and in expiration by decreasing the airway pressure according to the non-linear pressure-flow relationship of the ETT.
DESIGN: Prospective randomized single blind cross-over study.
SETTING: Laboratory of the Section of Experimental Anaesthesiology (Clinic of Anaesthesiology; University of Freiburg).
SUBJECTS: Ten healthy volunteers.
INTERVENTIONS: The subjects breathed spontaneously through an ETT of 7.5 mm i.d. Three different ventilatory modes, each with a PEEP of 5 cmH2O, were presented in random order using the Dräger Evita 2 ventilator with prototype software: (1) IPS (10 cmH2O, 1 s ramp), (2) inspiratory ATC (ATC-in), (3) inspiratory and expiratory ATC (ATC-in-ex).
MEASUREMENTS AND MAIN RESULTS: Immediately following a mode transition, the volunteers answered with a hand sign to show how they perceived the new mode compared with the preceding mode in terms of gain or loss in subjective respiratory comfort: "better", "unchanged" or "worse". Inspiration and expiration were investigated separately analyzing 60 mode transitions each. Flow rates were continuously measured. The transition from IPS to either type of ATC was perceived positively, i.e. as increased comfort, whereas the opposite transition from ATC to IPS was perceived negatively, i.e. as decreased comfort. The transition from ATC-in to ATC-in-ex was perceived positively whereas the opposite mode transition was perceived negatively in expiration only. Tidal volume was 1220 +/- 404 ml during IPS and 1017 +/- 362 ml during ATC. The inspiratory peak flow rate was 959 +/- 78 ml/s during IPS and 1048 +/- 197 ml/s during ATC.
CONCLUSIONS: ATC provides an increase in respiratory comfort compared with IPS. The predominant cause for respiratory discomfort in the IPS mode seems to be lung over-inflation.

Entities:  

Mesh:

Year:  1997        PMID: 9434916     DOI: 10.1007/s001340050467

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  21 in total

1.  Additional work of breathing imposed by endotracheal tubes, breathing circuits, and intensive care ventilators.

Authors:  A D Bersten; A J Rutten; A E Vedig; G A Skowronski
Journal:  Crit Care Med       Date:  1989-07       Impact factor: 7.598

2.  Effect of perception of mechanical loading on human respiratory pattern regulation.

Authors:  J A Daubenspeck; E S Rhodes
Journal:  J Appl Physiol (1985)       Date:  1995-07

3.  Continuous calculation of intratracheal pressure in tracheally intubated patients.

Authors:  J Guttmann; L Eberhard; B Fabry; W Bertschmann; G Wolff
Journal:  Anesthesiology       Date:  1993-09       Impact factor: 7.892

4.  Effect of inspiratory flow rate on respiratory sensation and pattern of breathing.

Authors:  H L Manning; E J Molinary; J C Leiter
Journal:  Am J Respir Crit Care Med       Date:  1995-03       Impact factor: 21.405

5.  Pulmonary mechanics in normal subjects following endotracheal intubation.

Authors:  T J Gal
Journal:  Anesthesiology       Date:  1980-01       Impact factor: 7.892

6.  Proportional assist ventilation, a new approach to ventilatory support. Theory.

Authors:  M Younes
Journal:  Am Rev Respir Dis       Date:  1992-01

7.  Proportional assist ventilation. Results of an initial clinical trial.

Authors:  M Younes; A Puddy; D Roberts; R B Light; A Quesada; K Taylor; L Oppenheimer; H Cramp
Journal:  Am Rev Respir Dis       Date:  1992-01

8.  Effect of inspiratory flow rate on respiratory output in normal subjects.

Authors:  A Puddy; M Younes
Journal:  Am Rev Respir Dis       Date:  1992-09

9.  Comparison of breathing comfort during weaning with two ventilatory modes.

Authors:  A R Knebel; S L Janson-Bjerklie; J D Malley; A G Wilson; J J Marini
Journal:  Am J Respir Crit Care Med       Date:  1994-01       Impact factor: 21.405

10.  Respiratory function during pressure support ventilation.

Authors:  N R MacIntyre
Journal:  Chest       Date:  1986-05       Impact factor: 9.410

View more
  6 in total

1.  Electronic extubation--is it worth trying?

Authors:  R Kuhlen; R Rossaint
Journal:  Intensive Care Med       Date:  1997-11       Impact factor: 17.440

2.  The Comparison of Automatic Tube Compensation (ATC) and T-piece During Weaning.

Authors:  Çiğdem Selek; Perihan Ergin Özcan; Günseli Orhun; Evren Şentürk; İbrahim Özkan Akıncı; Nahit Çakar
Journal:  Turk J Anaesthesiol Reanim       Date:  2014-01-06

3.  Methods of liberation from mechanical ventilation: Which one is best?

Authors:  Ling Liu
Journal:  Front Med (Lausanne)       Date:  2022-08-16

4.  Prediction of extubation outcome: a randomised, controlled trial with automatic tube compensation vs. pressure support ventilation.

Authors:  Jonathan Cohen; Maury Shapiro; Elad Grozovski; Ben Fox; Shaul Lev; Pierre Singer
Journal:  Crit Care       Date:  2009-02-23       Impact factor: 9.097

5.  Endotracheal tubes and imposed work of breathing: what should we do about it, if anything?

Authors:  Richard D Branson
Journal:  Crit Care       Date:  2003-08-28       Impact factor: 9.097

6.  Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief?

Authors:  Yoshiko Maeda; Yuji Fujino; Akinori Uchiyama; Nobuyuki Taenaka; Takashi Mashimo; Masaji Nishimura
Journal:  Crit Care       Date:  2003-08-14       Impact factor: 9.097

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.