Clinical evaluation of the Unfolder.

R Olson, R Cameron, T Hovis, J Hunkeler, R Lindstrom, R Steinert.   

Abstract

PURPOSE: To evaluate the clinical acceptability of the Unfolder for implantation of AMO PhacoFlex II (model SI-30NB or SI-40NB) intraocular lenses (IOLs).
SETTING: Six investigational sites in five states in the United States: Utah, Texas, Missouri, Minnesota, and Massachusetts.
METHODS: One hundred one patients from six investigational sites had implantation of an SI-30NB or SI-40NB IOL. Investigators evaluated the clinical acceptability of the Unfolder by comparing their experience with the injector with that using other implantation systems. Internal incision size (dimensions) was measured three times during surgery using a calibrated instrument. Data on safety and efficacy were collected.
RESULTS: The investigators rated the Unfolder as clinically acceptable. Mean internal incision size after phacoemulsification but before IOL implantation was 3.0 mm +/- 0.1 (SD) (range 2.7 to 3.2 mm). Average internal incision size after IOL implantation was 3.1 +/- 0.1 mm. There were no reports of haptic deformation, lens optic damage, or postoperative complications related to the Unfolder. No adverse events were reported.
CONCLUSIONS: Investigators found the Unfolder easy to use. It provided controlled lens implantation through incisions 2.7 to 3.2 mm, with post-IOL implantation incisions ranging from 2.9 to 3.3 mm.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9423912     DOI: 10.1016/s0886-3350(97)80119-8

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  The IOL flip: rescue for foldable lens implantation gone wrong.

Authors:  C Y W Khng; K-T Yeo
Journal:  Br J Ophthalmol       Date:  2003-05       Impact factor: 4.638

2.  Scanning electron microscopy study of different one-piece foldable acrylic intraocular lenses after injection through microincisional cataract surgery cartridges.

Authors:  Paolo Cecchini; Rossella D'Aloisio; Chiara De Giacinto; Gianluca Turco; Daniele Tognetto
Journal:  Int Ophthalmol       Date:  2019-10-17       Impact factor: 2.031

3.  Intraoperative cracking of the AMO Phacoflex-II Silicone (SI-40NB) intraocular lens while implanting with the Unfolder silver series system using hydroxypropylmethylcellulose.

Authors:  Rohit Saxena; Ankur Sinha; Sudarshan Khokhar; Munish Dhawan; Kiran Turuka
Journal:  Indian J Ophthalmol       Date:  2007 Sep-Oct       Impact factor: 1.848

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.