Literature DB >> 9367309

Fast inversion-recovery MR: the effect of hybrid RARE readout on the null points of fat and cerebrospinal fluid.

E R Melhem1, H Jara, H Shakir, T A Gagliano.   

Abstract

PURPOSE: To evaluate the effect of the hybrid RARE (rapid acquisition with relaxation enhancement) readout, commonly coupled to inversion-recovery pulse sequences, on the null inversiton time (TI) of fluid and fat using both phantoms and human volunteers.
METHODS: Two phantoms, simulating fat (phantom A) and cerebrospinal fluid (phantom B), respectively, were imaged using a fast inversion-recovery sequence that coupled an inversion-recovery preparation pulse to a hybrid RARE readout. At repetition times (TRs) ranging from 700 to 20,000, the TI necessary to null the signal from each phantom (null TI) was determined for an echo train length of 4, 6, 8, 10, 12, 14, 16, 18, and 20, respectively. Plots of null TI versus echo train length at different TRs were generated for both phantoms. Fast inversion-recovery MR imaging of the cervical spine and brain was performed in healthy volunteers. At a fixed TR and TI, the adequacy of signal suppression from bone marrow and cerebrospinal fluid was assessed as a function of echo train length.
RESULTS: There was a gradual decrease of null TI for both phantoms with echo train length. This decrease persisted at longer TRs for phantom B (T1 = 3175 +/- 70 milliseconds) than for phantom A (T1 = 218 +/- 5 milliseconds). In the human volunteers, there was a gradual loss of suppression of signal from bone marrow and cerebrospinal fluid, with changes in the hybrid RARE readout.
CONCLUSION: To optimize specific tissue suppression, radiologists implementing fast inversion-recovery MR imaging should be aware of the effects of the hybrid RARE readout on null TI.

Entities:  

Mesh:

Year:  1997        PMID: 9367309      PMCID: PMC8338443     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  3 in total

1.  Influence of imaging parameters on high-intensity cerebrospinal fluid artifacts in fast-FLAIR MR imaging.

Authors:  Hsiu-Mei Wu; David M Yousem; Hsiao-Wen Chung; Wan-Yuo Guo; Cheng-Yen Chang; Cheng-Yu Chen
Journal:  AJNR Am J Neuroradiol       Date:  2002-03       Impact factor: 3.825

2.  A quantitative postmortem MRI design sensitive to white matter hyperintensity differences and their relationship with underlying pathology.

Authors:  Melissa E Murray; Prashanthi Vemuri; Greg M Preboske; Matthew C Murphy; Katherine J Schweitzer; Joseph E Parisi; Clifford R Jack; Dennis W Dickson
Journal:  J Neuropathol Exp Neurol       Date:  2012-12       Impact factor: 3.685

3.  Toward real-time temperature monitoring in fat and aqueous tissue during magnetic resonance-guided high-intensity focused ultrasound using a three-dimensional proton resonance frequency T1 method.

Authors:  Mahamadou Diakite; Henrik Odéen; Nick Todd; Allison Payne; Dennis L Parker
Journal:  Magn Reson Med       Date:  2013-07-30       Impact factor: 4.668

  3 in total