S T Wang1, S Pizzolato, H P Demshar.
Abstract
We used ROC plots to evaluate the clinical performance of the Guthrie, Wallac, and Isolab assays for newborn phenylketonuria (PKU) screening and assessed the screening discriminatory power of these three assays by the area under the ROC plot, Youden's J index, and the likelihood ratio. The use of these plots not only allows us to pinpoint the exact cutoff value in screening, but also provides a direct comparison of these three different assays in clinical outcome performance. The optimum cutoff for the newborn PKU screening is a blood phenylalanine concentration of 0.30, 0.27, and 0.18 mmol/L for the Guthrie, Wallac, and Isolab assays, respectively. We conclude that the Wallac and Isolab kits, like the Guthrie assay, are suitable for newborn PKU screening.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9342001
Source DB: PubMed Journal: Clin Chem ISSN: 0009-9147 Impact factor: 8.327