Literature DB >> 9338885

An experimental study of the failure modes of the Gamma Locking Nail and AO Dynamic Hip Screw under static loading: a cadaveric study.

R C Haynes1, R G Pöll, A W Miles, R B Weston.   

Abstract

The sliding compression screw is widely regarded as the optimum treatment for intertrochanteric fractures of the femur, allowing bone fragments to impact until a bony support has been established across the fracture site. This study carried out biomechanical, cadaveric tests to establish the influence of direct static loading situations on the modes of failure of the Gamma Nail compared with the Dynamic Hip Screw (DHS). Clinical studies report DHS failures of lag screws cutting-out, bending of the lag screws and cortical screws pulling out causing plate loosening. Gamma Nail failures include lag screw cut-out or fractures of the femoral shaft around the distal locking screws or nail tip. In this study each failure mode has been isolated, to establish the loads to failure under various fracture configurations. The biomechanical results indicated that the intramedullary Gamma Locking Nail can be recommended over a standard DHS in cases of subtrochanteric fracture or conditions of very poor bone quality.

Entities:  

Mesh:

Year:  1997        PMID: 9338885     DOI: 10.1016/s1350-4533(97)00003-9

Source DB:  PubMed          Journal:  Med Eng Phys        ISSN: 1350-4533            Impact factor:   2.242


  12 in total

1.  Treatment of subtrochanteric fractures. A comparison of the Gamma nail and the dynamic hip screw: short-term outcome in 58 patients.

Authors:  I Saarenpää; T Heikkinen; P Jalovaara
Journal:  Int Orthop       Date:  2006-04-22       Impact factor: 3.075

2.  Is tip apex distance as important as we think? A biomechanical study examining optimal lag screw placement.

Authors:  Patrick Kane; Bryan Vopat; Wendell Heard; Nikhil Thakur; David Paller; Sarath Koruprolu; Christopher Born
Journal:  Clin Orthop Relat Res       Date:  2014-04-24       Impact factor: 4.176

3.  Cephalomedullary nail versus sliding hip screw for fixation of AO 31 A1/2 intertrochanteric femoral fracture: a 12-year comparison of failure, complications, and mortality.

Authors:  Casey S Whale; D Andrew Hulet; Michael J Beebe; David L Rothberg; Chong Zhang; Angela P Presson; Ami R Stuart; Erik N Kubiak
Journal:  Curr Orthop Pract       Date:  2016 Nov-Dec

4.  Evaluation of complications of three different types of proximal extra-articular femur fractures: differences in complications, age, sex and surviving rates.

Authors:  Andreas A Suckel; Klaus Dietz; Nikolaus Wuelker; Peter Helwig
Journal:  Int Orthop       Date:  2006-10-11       Impact factor: 3.075

5.  Biopolymer augmentation of the lag screw in the treatment of femoral neck fractures--a biomechanical in-vitro study.

Authors:  A Paech; E Wilde; A P Schulz; G Heinrichs; R Wendlandt; C Queitsch; B Kienast; Ch Jürgens
Journal:  Eur J Med Res       Date:  2010-04-08       Impact factor: 2.175

6.  Cephalomedullary nailing for reverse oblique intertrochanteric fractures 31A3 (AO/OTA).

Authors:  Emanuela Marsillo; Andrea Pintore; Giovanni Asparago; Francesco Oliva; Nicola Maffulli
Journal:  Orthop Rev (Pavia)       Date:  2022-10-13

7.  Unusual mode of failure of intertrochanteric femur fracture fixation.

Authors:  Kunal Roy; Pradip Nemade; Rishi Aggarwal; Nirmal Patil
Journal:  BMJ Case Rep       Date:  2015-11-12

8.  Static load test of the modified sliding hip screw: the DHS-AF®.

Authors:  Anderson Freitas; Hélio Ismael da Costa; Célio José Silva; Carlos Henrique da Costa Rangel
Journal:  Acta Ortop Bras       Date:  2013       Impact factor: 0.513

9.  Double axis cephalocondylic fixation of stable and unstable intertrochanteric fractures: early results in 60 cases with the veronail system.

Authors:  Alpha Kouzelis; A Kravvas; S Mylonas; D Giannikas; A Panagopoulos
Journal:  Open Orthop J       Date:  2014-04-24

10.  A review of tip apex distance in dynamic hip screw fixation of osteoporotic hip fractures.

Authors:  Imran Haruna Abdulkareem
Journal:  Niger Med J       Date:  2012-10