Detection of dicrotic notch in arterial pressure signals.

S A Hoeksel1, J R Jansen, J A Blom, J J Schreuder.   

Abstract

OBJECTIVE: A novel algorithm to detect the dicrotic notch in arterial pressure signals is proposed. Its performance is evaluated using both aortic and radial artery pressure signals, and its robustness to variations in design parameters is investigated.
METHODS: Most previously published dicrotic notch detection algorithms scan the arterial pressure waveform for the characteristic pressure change that is associated with the dicrotic notch. Aortic valves, however, are closed by the backwards motion of aortic blood volume. We developed an algorithm that uses arterial flow to detect the dicrotic notch in arterial pressure waveforms. Arterial flow is calculated from arterial pressure using simulation results with a three-element windkessel model. Aortic valve closure is detected after the systolic upstroke and at the minimum of the first negative dip in the calculated flow signal.
RESULTS: In 7 dogs ejection times were derived from a calculated aortic flow signal and from simultaneously measured aortic flow probe data. A total of 86 beats was analyzed; the difference in ejection times was -0.6 +/- 5.4 ms (means +/- SD). The algorithm was further evaluated using 6 second epochs of radial artery pressure data measured in 50 patients. Model simulations were carried out using both a linear windkessel model and a pressure and age dependent nonlinear windkessel model. Visual inspection by an experienced clinician confirmed that the algorithm correctly identified the dicrotic notch in 98% (49 of 50) of the patients using the linear model, and 96% (48 of 50) of the patients using the nonlinear model. The position of the dicrotic notch appeared to be less sensitive to variations in algorithm's design parameters when a nonlinear windkessel model was used.
CONCLUSIONS: The detection of the dicrotic notch in arterial pressure signals is facilitated by first calculating the arterial flow waveform from arterial pressure and a model of arterial afterload. The method is robust and reduces the problem of detecting a dubious point in a decreasing pressure signal to the detection of a well-defined minimum in a derived signal.

Entities:  

Mesh:

Year:  1997        PMID: 9338845     DOI: 10.1023/a:1007414906294

Source DB:  PubMed          Journal:  J Clin Monit        ISSN: 0748-1977


  23 in total

1.  Continuous cardiac output monitoring by pulse contour during cardiac surgery.

Authors:  J R Jansen; K H Wesseling; J J Settels; J J Schreuder
Journal:  Eur Heart J       Date:  1990-12       Impact factor: 29.983

2.  Evaluation of an extraaortic counterpulsation device in severe cardiac failure.

Authors:  J A Zelano; W Ko; R Lazzaro; W D Lazenby; S Topaz; W J Kolff; O W Isom; K H Krieger
Journal:  Ann Thorac Surg       Date:  1992-01       Impact factor: 4.330

3.  A nonlinear model of the arterial system incorporating a pressure-dependent compliance.

Authors:  J K Li; T Cui; G M Drzewiecki
Journal:  IEEE Trans Biomed Eng       Date:  1990-07       Impact factor: 4.538

4.  Assessment of Windkessel as a model of aortic input impedance.

Authors:  D Burkhoff; J Alexander; J Schipke
Journal:  Am J Physiol       Date:  1988-10

5.  Computer model analysis of the radial artery pressure waveform.

Authors:  H A Schwid; L A Taylor; N T Smith
Journal:  J Clin Monit       Date:  1987-10

6.  Processing of arterial pressure waves with a digital computer.

Authors:  C F Starmer; P A McHale; J C Greenfield
Journal:  Comput Biomed Res       Date:  1973-02

7.  "Wavefronts" in the aorta--implications for the mechanisms of left ventricular ejection and aortic valve closure.

Authors:  C J Jones; M Sugawara
Journal:  Cardiovasc Res       Date:  1993-11       Impact factor: 10.787

8.  bpshape wk4: a computer program that implements a physiological model for analyzing the shape of blood pressure waveforms.

Authors:  W C Ocasio; D R Rigney; K P Clark; R G Mark
Journal:  Comput Methods Programs Biomed       Date:  1993-04       Impact factor: 5.428

9.  A real-time pressure algorithm.

Authors:  P Kinias; H A Fozzard; M J Norusis
Journal:  Comput Biol Med       Date:  1981       Impact factor: 4.589

10.  Automated infusion of nitroglycerin to control arterial hypertension during cardiac surgery.

Authors:  S A Hoeksel; J J Schreuder; J A Blom; J G Maessen; O C Penn
Journal:  Intensive Care Med       Date:  1996-07       Impact factor: 17.440

View more
  13 in total

Review 1.  The arterial Windkessel.

Authors:  Nico Westerhof; Jan-Willem Lankhaar; Berend E Westerhof
Journal:  Med Biol Eng Comput       Date:  2008-06-10       Impact factor: 2.602

2.  Diastolic pressure ratio: new approach and validation vs. the instantaneous wave-free ratio.

Authors:  Nils P Johnson; Wenguang Li; Xi Chen; Barry Hennigan; Stuart Watkins; Colin Berry; William F Fearon; Keith G Oldroyd
Journal:  Eur Heart J       Date:  2019-08-14       Impact factor: 29.983

3.  Reliability of effective arterial elastance using peripheral arterial pressure as surrogate for left ventricular end-systolic pressure.

Authors:  Manuel Ignacio Monge Garcia; Zhongping Jian; Jos J Settels; Feras Hatib; Maurizio Cecconi; Michael R Pinsky
Journal:  J Clin Monit Comput       Date:  2018-12-14       Impact factor: 2.502

4.  Accurate measurement of the pulse wave delay with imaging photoplethysmography.

Authors:  Alexei A Kamshilin; Igor S Sidorov; Laura Babayan; Maxim A Volynsky; Rashid Giniatullin; Oleg V Mamontov
Journal:  Biomed Opt Express       Date:  2016-11-16       Impact factor: 3.732

5.  Performance of a real-time dicrotic notch detection and prediction algorithm in arrhythmic human aortic pressure signals.

Authors:  Andrea Donelli; Jos R C Jansen; Bas Hoeksel; Paolo Pedeferri; Ramzi Hanania; Jan Bovelander; Francesco Maisano; Alessandro Castiglioni; Alfieri Ottavio; Jan J Schreuder
Journal:  J Clin Monit Comput       Date:  2002 Apr-May       Impact factor: 2.502

6.  Algorithm for identifying and separating beats from arterial pulse records.

Authors:  Ernesto F Treo; Myriam C Herrera; Max E Valentinuzzi
Journal:  Biomed Eng Online       Date:  2005-08-11       Impact factor: 2.819

7.  Comparison of Transpulmonary Thermodilution and Calibrated Pulse Contour Analysis with Pulmonary Artery Thermodilution Cardiac Output Measurements in Anesthetized Dogs.

Authors:  N A Garofalo; F J Teixeira-Neto; J C Rodrigues; S A Cerejo; A J A Aguiar; D R Becerra-Velásquez
Journal:  J Vet Intern Med       Date:  2016-05-30       Impact factor: 3.333

8.  Use of the Kalman Filter for Aortic Pressure Waveform Noise Reduction.

Authors:  Frank Lam; Hsiang-Wei Lu; Chung-Che Wu; Zekeriya Aliyazicioglu; James S Kang
Journal:  Comput Math Methods Med       Date:  2017-05-22       Impact factor: 2.238

9.  Cuffless Blood Pressure Estimation Using Pressure Pulse Wave Signals.

Authors:  Zeng-Ding Liu; Ji-Kui Liu; Bo Wen; Qing-Yun He; Ye Li; Fen Miao
Journal:  Sensors (Basel)       Date:  2018-12-02       Impact factor: 3.576

10.  Cardiac output derived from left ventricular pressure during conductance catheter evaluations: an extended Modelflow method.

Authors:  Sergio Valsecchi; Giovanni B Perego; Jan J Schreuder; Federica Censi; Jos R C Jansen
Journal:  J Clin Monit Comput       Date:  2007-06-01       Impact factor: 1.977

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.