Literature DB >> 9330133

Prospective multicenter trial of a plate for dorsal fixation of distal radius fractures.

D Ring1, J B Jupiter, J Brennwald, U Büchler, H Hastings.   

Abstract

A new plate designed specifically to address complex wrist pathology was used for the internal fixation of 22 complex fractures of the distal radius in 22 patients in a prospective multicenter trial. The majority of fractures were group C2- and C3-type fractures according to the Comprehensive Classification of Fractures. No plate failures, loss of reduction, nonunions, or infections occurred. Within the average follow-up time of 14 months, the functional results (including an average motion of 76% and an average grip strength of 56% of the contralateral side) were comparable to those reported for similar fractures in previous investigations. Five patients had irritation of the tendons in the second dorsal compartment. This trial serves both as a verification of the safety and efficacy of this distal radius plate as well as a demonstration of its utility in the treatment of complex fractures of the distal radius.

Entities:  

Mesh:

Year:  1997        PMID: 9330133     DOI: 10.1016/S0363-5023(97)80069-X

Source DB:  PubMed          Journal:  J Hand Surg Am        ISSN: 0363-5023            Impact factor:   2.230


  35 in total

Review 1.  Treatment of osteoporotic distal radius fractures.

Authors:  David Ring; Jesse B Jupiter
Journal:  Osteoporos Int       Date:  2004-12-22       Impact factor: 4.507

2.  Complex fractures of the distal radius treated with angular stability plates.

Authors:  Marco Frattini; Giovanni Soncini; Maurizio Corradi; Bruno Panno; Silvio Tocco; Francesco Pogliacomi
Journal:  Chir Organi Mov       Date:  2009-07-21

3.  Relevant Insertion Site Anatomy of the Conventus Distal Radius System.

Authors:  Jacob Duncan; Marc Trzeciak
Journal:  Hand (N Y)       Date:  2017-03-30

4.  [Treatment of distal radius fractures. Percutaneous Kirschner-wires or palmar locking plates?].

Authors:  R Meier; C Krettek; C Probst
Journal:  Unfallchirurg       Date:  2012-07       Impact factor: 1.000

5.  Dorsal Lunate Facet Fracture Reduction Using a Bone Reduction Forceps.

Authors:  Jonathan Lans; Josefina Alvarez; Amir R Kachooei; Sezai Ozkan; Jesse B Jupiter
Journal:  J Wrist Surg       Date:  2018-10-03

6.  Scott's parabola and the rise of the medical-industrial complex.

Authors:  Jesse Jupiter; Dennis Burke
Journal:  Hand (N Y)       Date:  2013-09

Review 7.  Hand Surgeon Reporting of Tendon Rupture Following Distal Radius Volar Plating.

Authors:  Nathan A Monaco; C Liam Dwyer; Alex J Ferikes; John D Lubahn
Journal:  Hand (N Y)       Date:  2016-02-19

8.  Biomechanical comparison of different volar fracture fixation plates for distal radius fractures.

Authors:  Kareem Sobky; Todd Baldini; Kenneth Thomas; Joel Bach; Allison Williams; Jennifer Moriatis Wolf
Journal:  Hand (N Y)       Date:  2007-09-07

9.  Biomechanical evaluation of volar locking plates for distal radius fractures.

Authors:  Scott M Levin; Cory O Nelson; Jonathan D Botts; Glenn A Teplitz; Yong Kwon; Fred Serra-Hsu
Journal:  Hand (N Y)       Date:  2007-08-07

10.  Different osteosyntheses for Colles' fracture: a mechanical study in 42 cadaver bones.

Authors:  Yngvar Krukhaug; Nils R Gjerdet; Odd J Lundberg; Peer K Lilleng; Leiv M Hove
Journal:  Acta Orthop       Date:  2009-04       Impact factor: 3.717