Literature DB >> 9306287

Relationship between venous pressure and tissue volume during venous congestion plethysmography in man.

F Christ1, J Gamble, H Baschnegger, I B Gartside.   

Abstract

1. Venous congestion strain-gauge plethysmography enables the non-invasive assessment of arterial blood flow, fluid filtration capacity (Kf), venous pressure (Pv) and isovolumetric venous pressure (Pvi) in man. One of the major assumptions of this technique, that cuff pressure (Pcuff) applied to the limb equals Pv at the level of the strain gauge, was tested in this study. 2. In nine healthy male volunteers (mean age, 29.3 +/- 1.2 years) the saphenous vein was cannulated with an 18-gauge catheter proximal to the medial malleolus. The subjects were supine and Pv was continuously measured during the application of small step (8-10 mmHg) increases in congestion Pcuff (up to 70 mmHg). Pcuff, changes in limb circumference and Pv were recorded by computer for off-line analysis. Since the determination of Kf is influenced by the changes in plasma oncotic pressure, venous blood samples were obtained at the start of the study, when Pcuff was raised to 30 mmHg and again to 65 mmHg and 4 min after deflation of the cuff. 3. The relationship between Pv and Pcuff was linear over the range of 10-70 mmHg (n = 9, 69 measurements, slope 0.91, r = 0.97, P << 0.001). The non-invasively measured calf Pv, based on the intercept of the relationship between the vascular compliance component (Va) and Pcuff, was 8.0 +/- 0.4 mmHg, which was not significantly different from the corrected invasively measured Pv value of 8.8 +/- 0.3 mmHg (P = 0.08). 4. Venous blood lactate and haemoglobin concentrations, as well as colloid osmotic pressure, total protein and albumin concentrations were unchanged throughout the protocol, whereas significant decreases in PO2 and blood glucose concentration were observed when Pcuff reached 65 mmHg. Assuming a constant oxygen consumption, this may suggest a reduction in tissue perfusion. 5. This study demonstrates the close correlation between Pcuff and Pv in the saphenous vein. Since the small congestion Pcuff step protocol does not cause significant increase in plasma oncotic pressure, we conclude that Pv, as well as Kf, can be accurately determined with this venous congestion plethysmography protocol.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9306287      PMCID: PMC1159877          DOI: 10.1111/j.1469-7793.1997.463bh.x

Source DB:  PubMed          Journal:  J Physiol        ISSN: 0022-3751            Impact factor:   5.182


  13 in total

1.  VASCULAR ADJUSTMENTS TO INCREASED TRANSMURAL PRESSURE IN CAT AND MAN WITH SPECIAL REFERENCE TO SHIFTS IN CAPILLARY FLUID TRANSFER.

Authors:  S MELLANDER; B OBERG; H ODELRAM
Journal:  Acta Physiol Scand       Date:  1964 May-Jun

2.  THE EFFECT OF POSTURE (STANDING) ON THE SERUM PROTEIN CONCENTRATION AND COLLOID OSMOTIC PRESSURE OF BLOOD FROM THE FOOT IN RELATION TO THE FORMATION OF EDEMA.

Authors:  J B Youmans; H S Wells; D Donley; D G Miller; H Frank
Journal:  J Clin Invest       Date:  1934-05       Impact factor: 14.808

3.  Skeletal muscle oxygenation monitoring by near infrared spectroscopy.

Authors:  R A De Blasi; E Quaglia; M Ferrari
Journal:  Biochem Int       Date:  1991-09

4.  The effect of passive tilting on microvascular parameters in the human calf: a strain gauge plethysmography study.

Authors:  J Gamble; F Christ; I B Gartside
Journal:  J Physiol       Date:  1997-01-15       Impact factor: 5.182

5.  Deep venous contribution to hydrostatic blood volume change in the human leg.

Authors:  J C Buckey; R M Peshock; C G Blomqvist
Journal:  Am J Cardiol       Date:  1988-09-01       Impact factor: 2.778

Review 6.  Interstitial-lymphatic mechanisms in the control of extracellular fluid volume.

Authors:  K Aukland; R K Reed
Journal:  Physiol Rev       Date:  1993-01       Impact factor: 37.312

7.  Duration of increase in vascular volume during venous stasis.

Authors:  P Sejrsen; O Henriksen; W P Paaske; S L Nielsen
Journal:  Acta Physiol Scand       Date:  1981-03

8.  Filling and emptying of the low-pressure blood vessels of the human forearm.

Authors:  E Brown; D M Greenfield; J S Goei; G Plassaras
Journal:  J Appl Physiol       Date:  1966-03       Impact factor: 3.531

9.  Changes in subcutaneous interstitial fluid pressure, tissue oxygenation, and skin red cell flux during venous congestion plethysmography in men.

Authors:  F Christ; M Dellian; A E Goetz; J Gamble; K Messmer
Journal:  Microcirculation       Date:  1997-03       Impact factor: 2.628

10.  Plasma colloid osmotic pressure in venous blood from the human foot in orthostasis.

Authors:  H Noddeland; K Aukland; G Nicolaysen
Journal:  Acta Physiol Scand       Date:  1981-12
View more
  13 in total

Review 1.  Assessment of venous capacitance. Radionuclide plethysmography: methodology and research applications.

Authors:  Matthias Schmitt; Daniel J Blackman; Gordon W Middleton; John R Cockcroft; Michael P Frenneaux
Journal:  Br J Clin Pharmacol       Date:  2002-12       Impact factor: 4.335

2.  Influence of different cuff inflation protocols on capillary filtration capacity in human calves -- a congestion plethysmography study.

Authors:  A Bauer; D Bruegger; J Gamble; F Christ
Journal:  J Physiol       Date:  2002-09-15       Impact factor: 5.182

3.  Age- and limb-related differences in the vasoconstrictor response to limb dependency are not mediated by a sympathetic mechanism in humans.

Authors:  K A M Snyder; S Shamimi-Noori; T E Wilson; K D Monahan
Journal:  Acta Physiol (Oxf)       Date:  2012-02-18       Impact factor: 6.311

4.  Cutaneous postural vasoconstriction is modified by exogenous but not endogenous female hormones in young women.

Authors:  Gemma D Bishop; Margaret D Brown
Journal:  Clin Auton Res       Date:  2007-01-30       Impact factor: 4.435

5.  Human calf precapillary resistance decreases in response to small cumulative increases in venous congestion pressure.

Authors:  J Gamble; F Christ; I B Gartside
Journal:  J Physiol       Date:  1998-03-01       Impact factor: 5.182

6.  Increased plasma angiotensin II in postural tachycardia syndrome (POTS) is related to reduced blood flow and blood volume.

Authors:  Julian M Stewart; June L Glover; Marvin S Medow
Journal:  Clin Sci (Lond)       Date:  2006-02       Impact factor: 6.124

7.  Regional blood volume and peripheral blood flow in postural tachycardia syndrome.

Authors:  Julian M Stewart; Leslie D Montgomery
Journal:  Am J Physiol Heart Circ Physiol       Date:  2004-04-29       Impact factor: 4.733

8.  A double-blind placebo-controlled cross-over study of the vascular effects of midodrine in neuropathic compared with hyperadrenergic postural tachycardia syndrome.

Authors:  Amanda J Ross; Anthony J Ocon; Marvin S Medow; Julian M Stewart
Journal:  Clin Sci (Lond)       Date:  2014-02       Impact factor: 6.124

9.  Impact of tetraplegia vs. paraplegia on venoarteriolar, myogenic and maximal cutaneous vasodilation responses of the microvasculature: Implications for cardiovascular disease.

Authors:  Michelle Trbovich; Yubo Wu; Wouter Koek; Joan Zhao; Dean Kellogg
Journal:  J Spinal Cord Med       Date:  2020-06-04       Impact factor: 1.985

10.  A Novel Non-Invasive Device for the Assessment of Central Venous Pressure in Hospital, Office and Home.

Authors:  Emanuela Marcelli; Laura Cercenelli; Barbara Bortolani; Saverio Marini; Luca Arfilli; Alessandro Capucci; Gianni Plicchi
Journal:  Med Devices (Auckl)       Date:  2021-05-13
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.