[Clinical use of software-controlled x-ray tube modulation with "Smart-Scan" in spiral CT].

K J Lehmann1, J Wild, M Georgi.   

Abstract

PURPOSE: The modulation of tube current during 360 degree-rotation for equalization of local differences in absorption is able to reduce patient exposure in spiral-CT. The impact on CT investigations of thorax, abdomen and pelvis and the influence of patient geometry within the slice on the decrease of tube current are evaluated.
METHODS: 300 spiral-CT's were performed on a HiSpeed Advantage scanner (GE) using a software-triggered modulation of tube current ("Smart-Scan"). The local absorption was measured in the 100 central channels of 2 scoutiviews (lateral and a.p.) and used for anatomical adapted modulation. The reduction of exposure as well as the height and width of the body cross-section were measured in reference slices of 100 thorax, 100 abdomen and 100 pelvis examinations.
RESULTS: The use of "Smart-Scar" enables a decrease of tube current of 7% in thorax, 6% in abdomen and 11% pelvis. There is a significant correlation between the height and width of body cross-section and the reduction of exposure for examinations of thorax and abdomen, but not of the pelvis.
CONCLUSION: Independent of patient geometry the clinical use of "Smart-Scan" achieves the pelvis. Anatomical regions with an oval body cross-section show a decrease of x-ray tube current, too.

Entities:  

Mesh:

Year:  1997        PMID: 9296614

Source DB:  PubMed          Journal:  Aktuelle Radiol        ISSN: 0939-267X


  4 in total

1.  Left ventricle volume measurements in cardiac micro-CT: the impact of radiation dose and contrast agent.

Authors:  Cristian T Badea; Arthur W Wetzel; Nilesh Mistry; Stuart Pomerantz; Demian Nave; G Allan Johnson
Journal:  Comput Med Imaging Graph       Date:  2008-04       Impact factor: 4.790

2.  Dose reduction in subsecond multislice spiral CT examination of children by online tube current modulation.

Authors:  H Greess; J Lutze; A Nömayr; H Wolf; T Hothorn; W A Kalender; W Bautz
Journal:  Eur Radiol       Date:  2004-03-30       Impact factor: 5.315

3.  Low-dose cone-beam CT via raw counts domain low-signal correction schemes: Performance assessment and task-based parameter optimization (Part II. Task-based parameter optimization).

Authors:  Daniel Gomez-Cardona; John W Hayes; Ran Zhang; Ke Li; Juan Pablo Cruz-Bastida; Guang-Hong Chen
Journal:  Med Phys       Date:  2018-04-06       Impact factor: 4.071

4.  Evaluation of AAPM Reports 204 and 220: Estimation of effective diameter, water-equivalent diameter, and ellipticity ratios for chest, abdomen, pelvis, and head CT scans.

Authors:  Christiane S Burton; Timothy P Szczykutowicz
Journal:  J Appl Clin Med Phys       Date:  2017-11-27       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.