An experimental model for the acute and chronic evaluation of intra-aneurysmal pressure.

P L Faries1, L A Sanchez, M L Marin, R E Parsons, R T Lyon, S Oliveri, F J Veith.   

Abstract

PURPOSE: To develop an animal model for the acute and chronic monitoring of pressure within abdominal aortic aneurysms (AAAs) to be treated with endovascular grafts.
METHODS: A strain-gauge pressure transducer was placed within an AAA created from a prosthetic vascular graft. Prosthetic aneurysms were implanted into 17 canine infrarenal aortas. The intra-aneurysmal pressure was monitored and correlated with noninvasive forelimb sphygmomanometry for 2 weeks. After this time, an intravascular manometer catheter was passed into the aneurysm. Simultaneous pressure measurements were obtained using the implanted strain-gauge pressure transducer, the manometer catheter, and the forelimb sphygmomanometer. Angiography was performed to assess intraluminal morphology, aneurysm anastomoses, and adjoining aortic vessels. In addition, two control animals underwent intra-aneurysmal pressure monitoring after standard surgical aneurysm repair.
RESULTS: There was excellent correlation (r = 0.97) between the pressure measurements obtained with the implanted strain-gauge pressure transducer and the intravascular manometer. Close correlation was also observed between the implanted strain-gauge transducer and the forelimb sphygmomanometer (r = 0.88) during postprocedural monitoring. Intra-aneurysmal pressure was lowered dramatically by surgical exclusion (aneurysm: 15/5 +/- 7/4 mmHg; systemic: 124/66 +/- 34/17 mmHg; p < 0.001). The prosthetic aneurysms were successfully imaged with angiography.
CONCLUSIONS: This animal model provides an accurate and reproducible means for measuring intra-aneurysmal pressure on an acute and chronic basis. It may be possible to use this model in the assessment of endovascular devices to determine their efficacy in reducing intra-aneurysmal pressure. Evaluation of complications associated with their use, such as patent aneurysm side branches, perigraft channels, and perianastomotic reflux, may also be possible.

Entities:  

Mesh:

Year:  1997        PMID: 9291056     DOI: 10.1583/1074-6218(1997)004<0290:AEMFTA>2.0.CO;2

Source DB:  PubMed          Journal:  J Endovasc Surg        ISSN: 1074-6218


  3 in total

1.  Significance of initial aortic aneurysm pressure sensor readings varies with aortic endograft design.

Authors:  Gautam Shrikhande; Sikandar Khan; Hafiz Hussain; Alicia Menezes; James McKinsey; Harry Bush
Journal:  World J Surg       Date:  2010-12       Impact factor: 3.352

2.  Aortic aneurysm sac pressure measurements after endovascular repair using an implantable remote sensor: initial experience and short-term follow-up.

Authors:  Hanno Hoppe; Jocelyn A Segall; Timothy K Liem; Gregory J Landry; John A Kaufman
Journal:  Eur Radiol       Date:  2007-12-20       Impact factor: 5.315

3.  Correlation between intrasac pressure measurements of a pressure sensor and an angiographic catheter during endovascular repair of abdominal aortic aneurysm.

Authors:  Pierre Galvagni Silveira; Christopher William Teixeira Miller; Rafael Freygang Mendes; Gilberto Nascimento Galego
Journal:  Clinics (Sao Paulo)       Date:  2008-02       Impact factor: 2.365

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.