J Cygler1, X A Li, G X Ding, E Lawrence.
Abstract
This paper presents a practical approach to clinical dosimetry for electron beams at extended source-to-surface distance (SSD). Characteristics of electron beams from a Siemens MD-2 accelerator are presented at a nominal SSD of 100 cm and an extended SSD of 115 cm. Relative output factors are measured at both 100 and 115 cm SSDs for a range of square field sizes for each beam energy. The change of output with SSD does not follow the inverse square law, if the nominal SSD is used. This deviation is larger for lower beam energy and smaller field sizes. An effective SSD, SSDeff, has been determined at dmax for each beam energy as a function of field size. A comprehensive approach to treatment time calculations is proposed.Mesh:
Year: 1997 PMID: 9279902 DOI: 10.1088/0031-9155/42/8/003
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609