Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant.

R J Bigsby1, C S Hardaker, J Fisher.   

Abstract

The Leeds physiological anatomical (PA) hip joint simulator was developed to apply three axes of loading and a complex three-dimensional motion so that the forces and motions can reproduce exactly the walking cycles defined by Paul. This paper presents the results of a study using the Leeds PA hip joint simulator to determine the wear of 32 mm ultra-high molecular weight polyethylene (UHMWPE) acetabular cups against stainless steel and zirconia ceramic heads, using bovine serum as lubricant. These results have been compared with the results of a previous study that used water as the lubricant, which led to UHMWPE transfer film being formed on the stainless steel head. Comparisons are also made with clinical results and results from other simulators. The study indicates that it is preferable to use bovine serum in simulator studies. In addition, the results indicate that if the surface roughness of the metallic and femoral heads are similar, and they remain undamaged during the tests, the wear rates of the UHMWPE cups are likely to be similar.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9256003     DOI: 10.1243/0954411971534377

Source DB:  PubMed          Journal:  Proc Inst Mech Eng H        ISSN: 0954-4119            Impact factor:   1.617


  3 in total

1.  Effect of motion inputs on the wear prediction of artificial hip joints.

Authors:  Feng Liu; John Fisher; Zhongmin Jin
Journal:  Tribol Int       Date:  2013-07       Impact factor: 4.872

2.  Influence of hip joint simulator design and mechanics on the wear and creep of metal-on-polyethylene bearings.

Authors:  Murat Ali; Mazen Al-Hajjar; Susan Partridge; Sophie Williams; John Fisher; Louise M Jennings
Journal:  Proc Inst Mech Eng H       Date:  2016-05       Impact factor: 1.617

3.  (iv) Enhancing the safety and reliability of joint replacement implants.

Authors:  Louise M Jennings; Mazen Al-Hajjar; Claire L Brockett; Sophie Williams; Joanne L Tipper; Eileen Ingham; John Fisher
Journal:  Orthop Trauma       Date:  2012-08
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.