| Literature DB >> 9252043 |
K Thorré1, M Pravda, S Sarre, G Ebinger, Y Michotte.
Abstract
An automated microbore liquid chromatographic assay with dual electrochemical detection is described for the determination of serotonin, dopamine and their metabolites, 5-hydroxyindoleacetic acid, 3,4-dihydroxyphenylacetic acid and homovanillic acid. Due to the chemical instability of the compounds, the addition of an antioxidant is required for automated analysis over a long period of time (e.g., 20 h). Therefore, the time stability of these substances was tested with different antioxidants. The stability for serotonin and 5-hydroxyindoleacetic acid was poor in acidic medium containing Na2EDTA but could greatly be improved by the addition of L-cysteine and ascorbic acid. Using this assay, the neurotransmitters and their metabolites could easily be determined in microdialysates obtained from different rat brain areas.Entities:
Mesh:
Substances:
Year: 1997 PMID: 9252043 DOI: 10.1016/s0378-4347(97)00126-6
Source DB: PubMed Journal: J Chromatogr B Biomed Sci Appl ISSN: 1387-2273