Fetal dose for a patient undergoing mantle field irradiation for Hodgkin's disease.

J Cygler1, G X Ding, W Kendal, P Cross.   

Abstract

In order to safely treat a 23 weeks pregnant woman for supradiapharagmatic Hodgkin's disease, without compromising the fetus, a custom shielding table was constructed and extensive phantom measurements were performed. For 10 MV photons, the optimal shielding combination was found to consist of a 5 cm thick lead sheet placed onto a 1.25 cm aluminum supporting plate. The structure was placed directly above the phantom, over the region corresponding to the woman's abdomen, without any intervening air gap. By this means the dose to the fetus from machine leakage and collimator scatter was eliminated; the only remaining dose was due to in-phantom scatter. The woman was treated using a mantle field to a dose of 35 Gy in 20 fractions. The accumulated dose to the woman's uterine fundus and to her pubis were monitored with theroluminescent dosimeters. After completion of mantle therapy the doses to the fundus and pubis were 10 and 3 cGy, respectively. The fetal exposure was thus limited to below 10 cGy, within the zone of fetal tolerance. A normal infant was delivered at term.

Entities:  

Mesh:

Year:  1997        PMID: 9243468     DOI: 10.1016/s0958-3947(97)00011-3

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  6 in total

1.  Conceptus dose from involved-field radiotherapy for Hodgkin's lymphoma on a linear accelerator equipped with MLCs.

Authors:  Michalis Mazonakis; Efrossini Lyraraki; Charalambos Varveris; Elina Samara; Kyveli Zourari; John Damilakis
Journal:  Strahlenther Onkol       Date:  2009-06-09       Impact factor: 3.621

Review 2.  The Management of Lymphoma in the Setting of Pregnancy.

Authors:  Chelsea C Pinnix; Therese Y Andraos; Sarah Milgrom; Michelle A Fanale
Journal:  Curr Hematol Malig Rep       Date:  2017-06       Impact factor: 3.952

3.  A study of the shielding used to reduce leakage and scattered radiation to the fetus in a pregnant patient treated with a 6-MV external X-ray beam.

Authors:  Bin Han; Bryan Bednarz; X George Xu
Journal:  Health Phys       Date:  2009-12       Impact factor: 1.316

4.  A feasibility study to calculate unshielded fetal doses to pregnant patients in 6-MV photon treatments using Monte Carlo methods and anatomically realistic phantoms.

Authors:  Bryan Bednarz; X George Xu
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

5.  Monte Carlo Study of Fetal Dosimetry Parameters for 6 MV Photon Beam.

Authors:  Maryam Atarod; Parvaneh Shokrani
Journal:  J Med Signals Sens       Date:  2013-01

6.  Revisiting fetal dose during radiation therapy: evaluating treatment techniques and a custom shield.

Authors:  Amir M Owrangi; Donald A Roberts; Elizabeth L Covington; James A Hayman; Kathryn M Masi; Choonik Lee; Jean M Moran; Joann I Prisciandaro
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.