Literature DB >> 9225792

Quantification of systolic count increase in technetium-99m-MIBI gated myocardial SPECT.

K Fukuchi1, T Uehara, T Morozumi, E Tsujimura, S Hasegawa, K Yutani, H Kusuoka, T Nishimura.   

Abstract

UNLABELLED: This study was performed to clarify the validity of quantification of myocardial wall thickening by the count increase method using electrocardiography (ECG)-gated SPECT.
METHODS: We performed a phantom study to examine the quantification of this method and to clarify the relationship between the changes of relative counts and objective size (such as myocardial wall) under various conditions. In addition, in volunteers, the percent count increase (%CI) was analyzed in left ventricular segments based on circumferential profile curve analysis by ECG-gated SPECT with 99mTc-MIBI (methoxyisobutyl-isonitrile), and it was compared with the regional systolic wall thickness (%Th) assessed by echocardiography during low-dose dobutamine infusion.
RESULTS: In our phantom study, the relative count changes were correlated linearly with the object size only within less than 20 mm. Recovery coefficient curves were influenced by acquisition parameters such as type of collimator, diameter of camera rotation, counts and photon scattering. In ECG-gated SPECT, the %CI value was increased gradually at each stage after dobutamine infusion, in relation to the increase of the %Th seen on echocardiography, although there are significant large deviations between these two parameters.
CONCLUSIONS: In this study, quantitative analysis based on the %CI in ECG-gated SPECT may underestimate regional wall thickening. These data should be considered in the evaluation of the %CI as an index of myocardial function.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9225792

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  9 in total

1.  Two- and three-dimensional assessments of myocardial perfusion and function by using technetium-99m sestamibi gated SPECT with a combination of count- and image-based techniques.

Authors:  T Nakata; Y Katagiri; Y Odawara; M Eguchi; M Kuroda; K Tsuchihashi; M Hareyama; K Shimamoto
Journal:  J Nucl Cardiol       Date:  2000 Nov-Dec       Impact factor: 5.952

2.  Evaluation of left ventricular wall motion, volumes, and ejection fraction by gated myocardial tomography with technetium 99m-labeled tetrofosmin: a comparison with cine magnetic resonance imaging.

Authors:  P Vaduganathan; Z X He; G W Vick; J J Mahmarian; M S Verani
Journal:  J Nucl Cardiol       Date:  1999 Jan-Feb       Impact factor: 5.952

3.  Quantitative fluorine 18 deoxyglucose uptake by myocardial positron emission tomography in rats.

Authors:  Nobuhiro Handa; Yasuhiro Magata; Eiji Tadamura; Takahiro Mukai; Takeshi Nishina; Senri Miwa; Yutaka Sakakibara; Takuya Nomoto; Junji Konishi; Kazunobu Nishimura; Masashi Komeda
Journal:  J Nucl Cardiol       Date:  2002 Nov-Dec       Impact factor: 5.952

Review 4.  Gated SPECT in assessment of regional and global left ventricular function: major tool of modern nuclear imaging.

Authors:  Aiden Abidov; Guido Germano; Rory Hachamovitch; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2006 Mar-Apr       Impact factor: 5.952

Review 5.  Gated SPECT in assessment of regional and global left ventricular function: an update.

Authors:  Aiden Abidov; Guido Germano; Rory Hachamovitch; Piotr Slomka; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2013-12       Impact factor: 5.952

6.  Quantitative gated blood pool SPECT for the assessment of coronary artery disease at rest.

Authors:  M W Groch; R C Marshall; W D Erwin; D J Schippers; C A Barnett; E M Leidholdt
Journal:  J Nucl Cardiol       Date:  1998 Nov-Dec       Impact factor: 5.952

7.  Segmental and global left ventricular function assessment using gated SPECT with a semiconductor Cadmium Zinc Telluride (CZT) camera: phantom study and clinical validation vs cardiac magnetic resonance.

Authors:  Alban Bailliez; Tanguy Blaire; Frédéric Mouquet; R Legghe; B Etienne; Damien Legallois; Denis Agostini; Alain Manrique
Journal:  J Nucl Cardiol       Date:  2014-05-09       Impact factor: 5.952

8.  Do patient data ever exceed the partial volume limit in gated SPECT studies?

Authors:  K Nichols; E G DePuey; M I Friedman; A Rozanski
Journal:  J Nucl Cardiol       Date:  1998 Sep-Oct       Impact factor: 5.952

9.  Evaluation of left ventricular function and volume in patients with dilated cardiomyopathy: gated myocardial single-photon emission tomography (SPECT) versus echocardiography.

Authors:  Fatma Berk; Serkan Isgoren; Hakan Demir; Guliz Kozdag; Dilek Ural; Baki Komsuoglu
Journal:  Ann Saudi Med       Date:  2005 May-Jun       Impact factor: 1.526

  9 in total