Determination of a novel growth hormone secretagogue (MK-677) in human plasma at picogram levels by liquid chromatography with atmospheric pressure chemical ionization tandem mass spectrometry.

M L Constanzer1, C M Chavez-Eng, B K Matuszewski.   

Abstract

A sensitive and specific method for the determination of N-[1(R)¿[1,2-dihydro-1-methylsulfonylspiro(3H-indole-3,4'-piper idin)-1'-yl]carbonyl¿-2-(phenylmethoxy)ethyl]-2-amino-2-meth ylpropanamide (MK-677, I), a growth hormone secretagogue, has been developed. The method is based on high-performance liquid chromatography (HPLC) with tandem mass spectrometric (MS-MS) detection. The analyte and internal standard (II) were isolated from the basified plasma using a liquid-liquid extraction with methyl-tert.-butyl ether (MTBE). The organic extract was evaporated to dryness, the residue was reconstituted in mobile phase and injected into the HPLC system. The MS-MS detection was performed on a PE Sciex API III Plus tandem mass spectrometer using a heated nebulizer interface. Multiple reaction monitoring of parent-->product ion combinations at m/z 529-->267 and 527-->267 was used to quantify I and II, respectively. The assay was validated in human plasma in the concentration range of 0.1 to 100 ng/ml, and the limit of quantification (LOQ) was 0.1 ng/ml. The precision of the assay, as expressed as coefficients of variation (C.V.,%) was less than 7% at all concentrations within the standard curve range, with adequate assay specificity and accuracy. The HPLC-MS-MS method provided sufficient sensitivity to completely map the pharmacokinetic time-course following a single 5-mg oral dose of I.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9200526     DOI: 10.1016/s0378-4347(97)00035-2

Source DB:  PubMed          Journal:  J Chromatogr B Biomed Sci Appl        ISSN: 1387-2273


  2 in total

1.  Tandem mass spectrum of a growth hormone secretagogue: amide bond cleavage and resultant gas-phase rearrangement.

Authors:  Xue-Zhi Qin
Journal:  J Am Soc Mass Spectrom       Date:  2002-04       Impact factor: 3.109

2.  Automated nanospray using chip-based emitters for the quantitative analysis of pharmaceutical compounds.

Authors:  Leonard J Corkery; Henrianna Pang; Bradley B Schneider; Thomas R Covey; K W Michael Siu
Journal:  J Am Soc Mass Spectrom       Date:  2005-01-19       Impact factor: 3.109