[Artifacts in MR imaging of the temporomandibular joint caused by dental alloys: a phantom model study at T1.5].

C Fellner1, M Behr, F Fellner, P Held, G Handel, S Feuerbach.   

Abstract

PURPOSE: The influence of dental alloys on MRI of the temporomandibular joint was studied using a phantom model for this joint.
METHODS: At 1,5 T, 15 dental alloys and 14 of their most important components were investigated acquiring sagittal (FOV: 150 mm) and transverse (FOV: 250 mm) T1-weighted SE images. In 11 cases, T1- and T2*-weighted FLASH images were measured additionally. The artifacts were assessed qualitatively as well as quantitatively, and the samples were subdivided into four artifact categories.
RESULTS: Ag, Cu, Ga, In, Tl, Sn, Zn, amalgam, the precious alloys, the Au-Pd and Ag-Pd alloys showed no artifacts (category I). Minimal artifacts below 10 mm on transverse images (category II) were found for Cr, Pd, Pt and for the Ni-Cr alloy. Mn and the remaining non-precious alloys induced artifacts up to 30 mm (category III). Significant artifacts-more than 30 mm-(category IV) were caused by Ni-Cr and 18/8 wires and by Co, Fe, and Ni. T2*-weighted FLASH proved to be more susceptible for artifacts than T1-weighted SE and FLASH techniques.
CONCLUSIONS: In contrast to dental alloys for fixed prosthodontics, Ni-Cr- or 18/8 wires used for orthodontic bands can influence not only the image quality, but also the diagnostic reliability of MRI of the temporomandibular joint.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9198515     DOI: 10.1055/s-2007-1015452

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  3 in total

1.  The interobserver-validated relevance of intervertebral spacer materials in MRI artifacting.

Authors:  T Ernstberger; G Heidrich; T Bruening; S Krefft; G Buchhorn; H M Klinger
Journal:  Eur Spine J       Date:  2006-02-07       Impact factor: 3.134

2.  Implant detectibility of intervertebral disc spacers in post fusion MRI: evaluation of the MRI scan quality by using a scoring system--an in vitro study.

Authors:  Thorsten Ernstberger; Gabert Heidrich; Wolfgang Schultz; Eckhardt Grabbe
Journal:  Neuroradiology       Date:  2006-11-04       Impact factor: 2.804

3.  Metal artifact reduction in patients with dental implants using multispectral three-dimensional data acquisition for hybrid PET/MRI.

Authors:  Jeanne M Gunzinger; Gaspar Delso; Andreas Boss; Miguel Porto; Helen Davison; Gustav K von Schulthess; Martin Huellner; Paul Stolzmann; Patrick Veit-Haibach; Irene A Burger
Journal:  EJNMMI Phys       Date:  2014-12-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.