Literature DB >> 9187188

Pressure vs flow triggering during pressure support ventilation.

R Goulet1, D Hess, R M Kacmarek.   

Abstract

BACKGROUND: Adult mechanical ventilators have traditionally been pressure- or time-triggered. More recently, flow triggering has become available and some adult ventilators allow the choice between pressure or flow triggering. Prior studies have supported the superiority of flow triggering during continuous positive airway pressure, but few have compared pressure and flow triggering during pressure support ventilation (PSV). The purpose of this study was to compare pressure and flow triggering during PSV in adult mechanically ventilated patients.
METHODS: The study population consisted of 10 adult patients ventilated with a mechanical ventilator (Nellcor-Puritan-Bennett 7200ae) in the PSV mode. In random order, we compared pressure triggering of -0.5 H2O, pressure triggering -1 cm H2O, flow triggering of 5/2 L/min, and flow triggering 10/3 L/min. Pressure was measured for 5 min at the proximal endotracheal tube using a data acquisition rate of 100 Hz. From the airway pressure signal, trigger pressure (deltaP) was defined as the difference between positive end-expiratory pressure (PEEP) and the maximum negative deflection prior to onset of the triggered breath. Pressure-time product (PTP) was defined as the area produced by the pressure waveform below PEEP during onset of the triggered breath. Trigger time (deltaT) was defined as the time interval below PEEP during onset of the triggered breath.
RESULTS: A pressure trigger of -0.5 cm H2O was significantly more sensitive than the other trigger methods for deltaP, PTP, and deltaT (p<0.001). There was also a significant difference between patients for deltaP, deltaT, and PTP for each trigger method (p<0.001).
CONCLUSIONS: For this group of patients, flow triggering was not superior to pressure triggering at -0.5 cm H2O during PSV.

Entities:  

Mesh:

Year:  1997        PMID: 9187188     DOI: 10.1378/chest.111.6.1649

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  6 in total

1.  Neurally adjusted ventilator assist in very low birth weight infants: Current status.

Authors:  Hassib Narchi; Fares Chedid
Journal:  World J Methodol       Date:  2015-06-26

2.  Computational Modeling of a Low-Cost Fluidic Oscillator for Use in an Educational Respiratory Simulator.

Authors:  Tom Dillon; Caglar Ozturk; Keegan Mendez; Luca Rosalia; Samuel Dutra Gollob; Katharina Kempf; Ellen Tunney Roche
Journal:  Adv Nanobiomed Res       Date:  2021-11-14

3.  The effect of back-up rate during non-invasive ventilation in young patients with cystic fibrosis.

Authors:  Brigitte Fauroux; Bruno Louis; Nicholas Hart; Sandrine Essouri; Karl Leroux; Annick Clément; Michael Ian Polkey; Frédéric Lofaso
Journal:  Intensive Care Med       Date:  2004-01-16       Impact factor: 17.440

4.  Subject-ventilator synchrony during neural versus pneumatically triggered non-invasive helmet ventilation.

Authors:  Onnen Moerer; Jennifer Beck; Lukas Brander; Roberta Costa; Michael Quintel; Arthur S Slutsky; Fabrice Brunet; Christer Sinderby
Journal:  Intensive Care Med       Date:  2008-05-30       Impact factor: 17.440

5.  Neurally adjusted ventilatory assist and proportional assist ventilation both improve patient-ventilator interaction.

Authors:  Matthieu Schmidt; Felix Kindler; Jérôme Cecchini; Tymothée Poitou; Elise Morawiec; Romain Persichini; Thomas Similowski; Alexandre Demoule
Journal:  Crit Care       Date:  2015-02-25       Impact factor: 9.097

Review 6.  Patient-Ventilator Dyssynchrony.

Authors:  Elvira-Markela Antonogiannaki; Dimitris Georgopoulos; Evangelia Akoumianaki
Journal:  Korean J Crit Care Med       Date:  2017-11-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.