Literature DB >> 9186896

Hemodialyzer mass transfer-area coefficients for urea increase at high dialysate flow rates. The Hemodialysis (HEMO) Study.

J K Leypoldt1, A K Cheung, L Y Agodoa, J T Daugirdas, T Greene, P R Keshaviah.   

Abstract

The dialyzer mass transfer-area coefficient (KoA) for area is an important determinant of urea removal during hemodialysis and is considered to be constant for a given dialyzer. We determined urea clearance for 22 different models of commercial hollow fiber dialyzers (N = approximately 5/model, total N = 107) in vitro at 37 degrees C for three countercurrent blood (Qb) and dialysate (Qd) flow rate combinations. A standard bicarbonate dialysis solution was used in both the blood and dialysate flow pathways, and clearances were calculated from urea concentrations in the input and output flows on both the blood and dialysate sides. Urea KoA values, calculated from the mean of the blood and dialysate side clearances, varied between 520 and 1230 ml/min depending on the dialyzer model, but the effect of blood and dialysate flow rate on urea KoA was similar for each. Urea KoA did not change (690 +/- 160 vs. 680 +/- 140 ml/min, P = NS) when Qh increased from 306 +/- 7 to 459 +/- 10 ml/min at a nominal Qd of 500 ml/min. When Qd increased from 504 +/- 6 to 819 +/- 8 ml/min at a nominal Qh of 450 ml/min, however, urea KoA increased (P < 0.001) by 14 +/- 7% (range 3 to 33%, depending on the dialyzer model) to 780 +/- 150 ml/min. These data demonstrate that increasing nominal Qd from 500 to 800 ml/min alters the mass transfer characteristics of hollow fiber hemodialyzers and results in a larger increase in area clearance than predicted assuming a constant KoA.

Entities:  

Mesh:

Substances:

Year:  1997        PMID: 9186896     DOI: 10.1038/ki.1997.274

Source DB:  PubMed          Journal:  Kidney Int        ISSN: 0085-2538            Impact factor:   10.612


  14 in total

Review 1.  Extracorporeal Removal of Poisons and Toxins.

Authors:  Joshua David King; Moritz H Kern; Bernard G Jaar
Journal:  Clin J Am Soc Nephrol       Date:  2019-08-22       Impact factor: 8.237

2.  Characterization of the cylindrical electrospun nanofibrous polysulfone membrane for hemodialysis with modelling approach.

Authors:  Farideh Mohammadi; Farzaneh Mohammadi; Zeynab Yavari
Journal:  Med Biol Eng Comput       Date:  2021-07-17       Impact factor: 2.602

3.  Dialysate flow rate and delivered Kt/Vurea for dialyzers with enhanced dialysate flow distribution.

Authors:  Richard A Ward; John W Idoux; Hiba Hamdan; Rosemary Ouseph; Thomas A Depner; Thomas A Golper
Journal:  Clin J Am Soc Nephrol       Date:  2011-07-28       Impact factor: 8.237

4.  Pharmacometric characterization of dabigatran hemodialysis.

Authors:  Karl-Heinz Liesenfeld; Alexander Staab; Sebastian Härtter; Stephan Formella; Andreas Clemens; Thorsten Lehr
Journal:  Clin Pharmacokinet       Date:  2013-06       Impact factor: 6.447

5.  Dialysis dose and intradialytic hypotension: results from the HEMO study.

Authors:  Finnian R Mc Causland; Steven M Brunelli; Sushrut S Waikar
Journal:  Am J Nephrol       Date:  2013-10-26       Impact factor: 3.754

6.  Second Generation Nanoporous Silicon Nitride Membranes for High Toxin Clearance and Small Format Hemodialysis.

Authors:  Kayli Hill; Samuel N Walker; Alec Salminen; Hung L Chung; Xunzhi Li; Bahie Ezzat; Joshua J Miller; Jon-Paul S DesOrmeaux; Jingkai Zhang; Andrew Hayden; Tucker Burgin; Lindsay Piraino; Marina N May; Thomas R Gaborski; James A Roussie; Jeremy Taylor; Louis DiVincenti; Alexander A Shestopalov; James L McGrath; Dean G Johnson
Journal:  Adv Healthc Mater       Date:  2020-01-15       Impact factor: 9.933

7.  Effect of increasing dialysate flow rate on diffusive mass transfer of urea, phosphate and beta2-microglobulin during clinical haemodialysis.

Authors:  Jai P Bhimani; Rosemary Ouseph; Richard A Ward
Journal:  Nephrol Dial Transplant       Date:  2010-06-13       Impact factor: 5.992

Review 8.  Mass Transport in High-Flux Hemodialysis: Application of Engineering Principles to Clinical Prescription.

Authors:  Farzad Mohajerani; William R Clark; Claudio Ronco; Vivek Narsimhan
Journal:  Clin J Am Soc Nephrol       Date:  2022-03-11       Impact factor: 10.614

Review 9.  The Nitty-Gritties of Kt/Vurea Calculations in Hemodialysis and Peritoneal Dialysis.

Authors:  Brian Mark Churchill; Pallavi Patri
Journal:  Indian J Nephrol       Date:  2021-04-02

10.  Effects of high versus low flux membranes on o2 saturation in hemodialysis patients.

Authors:  Ali Momeni; Hamid Rouhi; Masoud Amiri
Journal:  Nephrourol Mon       Date:  2012-12-15
View more