| Literature DB >> 9174928 |
S Kossek1, C Padeste, L Tiefenauer.
Abstract
A modular immunosensor device based on silicon technology with electrochemical detection is presented. The sensor surface is functionalized with the antigen estradiol using the streptavidin-biotin system. The immobilized antigen is recognized by ferrocene-labeled antibody molecules in a competitive assay format. The ferrocene redox centers can be detected electrochemically. In order to facilitate the electron transfer, a gold layer structured in nanometer dimensions is used as an electrode. The functionalized silicon chip will be integrated into a micromachined fluid transport system. The resulting immunosensor device for estradiol represents a prototype allowing easy adaptation to other biotinylated analytes since the immobilized streptavidin serves as a universal anchor.Entities:
Mesh:
Substances:
Year: 1996 PMID: 9174928 DOI: 10.1002/(sici)1099-1352(199634/12)9:5/6<485::aid-jmr288>3.0.co;2-r
Source DB: PubMed Journal: J Mol Recognit ISSN: 0952-3499 Impact factor: 2.137