Perceptibility of defects in an aluminum test object: a comparison of the RVG-S and first generation VIXA systems with and without added niobium filtration.

M Wakoh1, A G Farman, W C Scarfe, M S Kelly, K Kuroyanagi.   

Abstract

OBJECTIVE: To compare the first generation VIXA (Gendex, Milan, Italy) and the RVG-S (Trophy Radiologie, Vincennes, France) for detection of defects in an aluminium test object at various exposures both with and without added niobium filtration.
METHODS: Images of a 7 mm aluminium test object with defects ranging from 0.1 to 1.5 mm with standardized projection geometry. Seven dentists acted as observers. Perceptibility curves were developed for both sensors under the various filtration conditions and exposures.
RESULTS: At optimum exposures both the VIXA and the RVG-S permitted the same number of defects to be observed. The optimum exposures were similar for the two systems, but the dynamic range was greater for the RVG-S. Addition of niobium filtration did not appreciably alter the radiation dose required to perceive a given number of defects using either system.
CONCLUSIONS: The two systems performed equally at optimum exposure when the dose is approximately the same irrespective of the presence or absence of niobium filtration. Beyond the optimal level, the RVG-S outperformed the VIXA system.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 9161163     DOI: 10.1259/dmfr.24.4.9161163

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  1 in total

1.  Comparison of psychophysical properties of two intraoral digital sensors on low-contrast perceptibility.

Authors:  X-Q Shi; D Benchimol; K Näsström
Journal:  Dentomaxillofac Radiol       Date:  2013-10-29       Impact factor: 2.419

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.