Vitek GPS card susceptibility testing accuracy using direct inoculation from BACTEC 9240 blood culture bottles.

W J Howard1, B J Buschelman, M J Bale, M A Pfaller, F P Koontz, R N Jones.   

Abstract

The emergent need for antimicrobial susceptibility testing (AST) data for the therapy of bacteremic patients has led to the development of rapid methods and local procedure modification of some commercial AST products such as the direct inoculation from blood culture systems. We compared the Vitek GPS card results using direct and standardized inoculation with a reference broth microdilution method for 112 consecutive staphylococcal bloodstream infections (seven drugs). Among the 28 Staphylococcus aureus strains, 0%-3.6% total error/drug was observed with both Vitek inoculation procedures. However, the only oxacillin-resistant strain was not detected (100% true very-major error). For 84 coagulase-negative staphylococci (CNS), the direct inoculation procedure had an 11.9% very-major error rate for oxacillin, ampicillin-sulbactam, and cephalothin, plus 4.8% very-major error rate for ciprofloxacin and trimethoprim-sulfamethoxazole (total error rate 4.8%-16.7% for five of seven drugs compared). The Vitek direct inoculation procedure routinely missed 20.4% of oxacillin-resistant CNS strains. The use of Vitek direct inoculation procedures for staphylococcal bloodstream infection isolates (from BACTEC 9240 cultures) produced serious false-susceptible results; this procedure should be avoided in favor of routine package insert-recommended Vitek procedures or other reference-quality overnight incubation susceptibility tests.

Entities:  

Mesh:

Year:  1996        PMID: 9147906     DOI: 10.1016/0732-8893(95)00273-1

Source DB:  PubMed          Journal:  Diagn Microbiol Infect Dis        ISSN: 0732-8893            Impact factor:   2.803


  7 in total

1.  Direct susceptibility testing of positive blood cultures by using Sensititre broth microdilution plates.

Authors:  Kimberle C Chapin; Michael C Musgnug
Journal:  J Clin Microbiol       Date:  2003-10       Impact factor: 5.948

2.  Accuracy of identification and susceptibility results by direct inoculation of Vitek 2 cards from positive BACTEC cultures.

Authors:  J J Kerremans; W H F Goessens; H A Verbrugh; M C Vos
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2004-12       Impact factor: 3.267

3.  Evaluation of the Merlin MICRONAUT system for rapid direct susceptibility testing of gram-positive cocci and gram-negative bacilli from positive blood cultures.

Authors:  Nele Wellinghausen; Tim Pietzcker; Sven Poppert; Syron Belak; Nicole Fieser; Melanie Bartel; Andreas Essig
Journal:  J Clin Microbiol       Date:  2007-01-03       Impact factor: 5.948

4.  Biographical Feature: Franklin P. Koontz, Ph.D., D(ABMM), F(AAM).

Authors:  Erik Munson
Journal:  J Clin Microbiol       Date:  2020-02-24       Impact factor: 5.948

5.  Oxacillin susceptibility testing of staphylococci directly from Bactec Plus blood cultures by the BBL Crystal MRSA ID system.

Authors:  M Kubina; B Jaulhac; X Delabranche; C Lindenmann; Y Piemont; H Monteil
Journal:  J Clin Microbiol       Date:  1999-06       Impact factor: 5.948

6.  Identification and susceptibility testing of Enterobacteriaceae and Pseudomonas aeruginosa by direct inoculation from positive BACTEC blood culture bottles into Vitek 2.

Authors:  Marjan J Bruins; Peter Bloembergen; Gijs J H M Ruijs; Maurice J H M Wolfhagen
Journal:  J Clin Microbiol       Date:  2004-01       Impact factor: 5.948

7.  Comparison of novel rapid diagnostic of blood culture identification and antimicrobial susceptibility testing by Accelerate Pheno system and BioFire FilmArray Blood Culture Identification and BioFire FilmArray Blood Culture Identification 2 panels.

Authors:  Dorothy T T Sze; Candy C Y Lau; Tsz-Ming Chan; Edmond S K Ma; Bone S F Tang
Journal:  BMC Microbiol       Date:  2021-12-18       Impact factor: 3.605

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.