Experimental correlation-based identification of X-ray CT point spread function. Part 2: Simulation and design of input signal.

S Doré1, R E Kearney, J A De Guise.   

Abstract

The preferred signals for non-parametric correlation-based point spread function identification are white noise or pseudo-random binary sequences (PRBSs). Given the difficulty of building a phantom based on either of these signals, a new input is devised that corresponds to pseudo-randomly located holes. The positions of the holes correspond to zeros in a 2-D PRBS. To optimise the design of the phantom and to ensure proper imaging procedure, a number of simulations are conducted. The effects of the following parameters on identification quality are investigated: the size of the holes and their minimum separation, the period of the PRBS, input-output translational and rotational mis-registration, pixel size and the presence of cupping. The factors affecting identification quality the most are rotational alignment, hole size and separation, as well as sequence length. During simulations, a point spread function offering characteristics similar to the Philips Tomoscan CX is identified. Optimal results are obtained when the signal consists of 0.6 mm holes, separated by 0.9 mm, whose position is based on a 32 x 32 PRBS generated with a ten-stage shift-register. When adequate rotational alignment is provided, it is shown that the pseudo-randomly located holes signal is a good substitute for a purely white signal when identifying the PSF of a CT scanner.

Entities:  

Mesh:

Year:  1997        PMID: 9136184     DOI: 10.1007/bf02510385

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  5 in total

1.  Experimental correlation-based identification of X-ray CT point spread function. Part 1: Method and experimental results.

Authors:  S Doré; R E Kearney; J A De Guise
Journal:  Med Biol Eng Comput       Date:  1997-01       Impact factor: 2.602

2.  The effects of misregistration of the projections on spatial resolution of CT scanners.

Authors:  M F Kijewski; P F Judy
Journal:  Med Phys       Date:  1983 Mar-Apr       Impact factor: 4.071

3.  The effects of sampling on CT images.

Authors:  P M Joseph; R D Spital; C D Stockham
Journal:  Comput Tomogr       Date:  1980 Jul-Sep

4.  Blurring in tomograms made with x-ray beams of finite width.

Authors:  J G Verly; R N Bracewell
Journal:  J Comput Assist Tomogr       Date:  1979-10       Impact factor: 1.826

5.  Correction for collimator width (restoration) in reconstructive X-ray tomography.

Authors:  R N Bracewell
Journal:  J Comput Assist Tomogr       Date:  1977-01       Impact factor: 1.826

  5 in total
  2 in total

1.  Experimental evaluation of computerised tomography point spread function variability within the field of view: parametric models.

Authors:  S Doré; R E Kearney
Journal:  Med Biol Eng Comput       Date:  2004-09       Impact factor: 2.602

2.  Experimental correlation-based identification of X-ray CT point spread function. Part 1: Method and experimental results.

Authors:  S Doré; R E Kearney; J A De Guise
Journal:  Med Biol Eng Comput       Date:  1997-01       Impact factor: 2.602

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.